6.4
No Heating or Insufficient Heating
Check if unit is pumping properly. If heat light is not on, check the setting of the setpoint temperature
and bath temperature. Check fluid level of the bath. Check for proper line voltage. Check for excessive
external cooling lead on unit. At higher temperatures, it could be due to heat loss from tanks, hoses or
vapor from a tank. Variance of heat load on the system from experiments may exceed the power
handling of the unit. Changes in heat load or setpoint requires time to settle to a stable temperature.
6.5 Triac Failure
Heater triac has failed or line supply voltage has a source of extreme interference from other equipment.
Try unit on another power source. If it still displays triac failure, a triac or triac driver needs replacement.
6.6 Default Settings
You may easily reset all parameters to factory default settings.
With unit in standby mode press the then then then select default EEPROM
Section 7. After-sale Support
All instruments requiring warranty repair must be returned to Brookfield Engineering Laboratories, Inc. or
the Brookfield dealer from whom it was purchased. Transportation is at the purchaser's expense.
For repair or service in the
United States, return to:
For repair or service
outside the United States, consult Brookfield Engineering Laboratories, Inc.
or the dealer from whom you purchased the instrument.
For repair or service in the
United Kingdom, return to :
Brookfield Viscometers Limited
1 Whitehall Estate Flex Meadow
Pinnacles West
Harlow, Essex CM19 5TJ, United Kingdom
Telephone (44) 27/945-1774 Fax: (44) 27/945-1775
For repair or service in
Germany, return to:
Brookfield Engineering Labs. Vertriebs GmbH
Barbarossastrasse 3
D-73547 Lorch, Germany
Telephone: 7172/927100
Fax: 7172/927105
Brookfield Engineering Labs., Inc.
11 Commerce Boulevard
Middleboro, MA 02346 USA
Telephone: 508-946-6200
Fax: 508-946-6262
Brookfield Viscometers Limited
1 Whitehall Estate
Flex Meadow
Pinnacles West
Harlow, Essex CM19 5TJ, United Kingdom
Telephone (44) 27/945-1774
Fax: (44) 27/945-1775
Brookfield Engineering Labs. Vertriebs GmbH
Barbarossastrasse 3
D-73547 Lorch, Germany
Telephone: 7172/927100
Fax: 7172/927105
[email protected]