VISCOMETER

TORQUE CONSTANT

MODEL CODE

MODEL

TK

ON DV-E SCREEN

LVDV-E

0.09373

LV

RVDV-E

1

RV

HADV-E

2

HA

HBDV-E

8

HB

SPECIAL ORDER TORQUE SPRINGS

VISCOMETER

TORQUE CONSTANT

MODEL CODE

MODEL

TK

ON DV-E SCREEN

2.5xLVDV-E

0.2343

2.5LV

5xLVDV-E

0.4686

5LV

1/4 RVDV-E

0.25

1/4RV

1/2 RVDV-E

0.5

1/2RV

2xHADV-E

4

2HA

2.5xHADV-E

5

2.5HA

2xHBDV-E

16

2HB

2.5xHBDV-E

20

2.5HB

Table C-2

This equation can be used to calculate the maximum viscosity that can be measured when using a

specific speed/spindle combination.

Full Scale Viscosity Range (FSR) = TK * SMC * 10,000

RPM