Brookfield DV-E Viscometer User Manual
Page 23
Brookfield Engineering Labs., Inc.
Page 23
Manual No. M98-350-J0912
VISCOMETER
TORQUE CONSTANT
MODEL CODE
MODEL
TK
ON DV-E SCREEN
LVDV-E
0.09373
LV
RVDV-E
1
RV
HADV-E
2
HA
HBDV-E
8
HB
SPECIAL ORDER TORQUE SPRINGS
VISCOMETER
TORQUE CONSTANT
MODEL CODE
MODEL
TK
ON DV-E SCREEN
2.5xLVDV-E
0.2343
2.5LV
5xLVDV-E
0.4686
5LV
1/4 RVDV-E
0.25
1/4RV
1/2 RVDV-E
0.5
1/2RV
2xHADV-E
4
2HA
2.5xHADV-E
5
2.5HA
2xHBDV-E
16
2HB
2.5xHBDV-E
20
2.5HB
Table C-2
This equation can be used to calculate the maximum viscosity that can be measured when using a
specific speed/spindle combination.
Full Scale Viscosity Range (FSR) = TK * SMC * 10,000
RPM
The units for FSR are in centipoise (cP). An example is shown in Appendix D.