beautypg.com

Brookfield DV-E Viscometer User Manual

Page 11

background image

Brookfield Engineering Labs., Inc.

Page 11

Manual No. M98-350-J0912

I.8 Cleaning

Make sure the instrument is in a decent working environment (dust-free, moderate
temperature, low humidity, etc.).

Make sure the instrument is on a level surface.

Hands/fingers must be clean and free of residual sample. Not doing so may result
in deposit build up on the upper part of the shaft and caue interference between
the shaft and the pivot cup.

Be sure to remove the spindle from the instrument prior to cleaning. Note left-
hand thread. Severe instrument damage may result if cleaned in place.

Instrument and Keypad:

Clean with dry, non-abrasive cloth. Do not use solvents
or cleaners.

Immersed Components (spindles): Spindles are made of stainless steel. Clean with non-

abrasive cloth and solvent appropriate for sample material

that is not aggressive to immersed components.

When cleaning, do not apply excessive force which may result in bending spindles.