beautypg.com

Small sample adapter (ssa) and thermosel (tsel) – Brookfield DV-E Viscometer User Manual

Page 17

background image

Brookfield Engineering Labs., Inc.

Page 17

Manual No. M98-350-J0912

Appendix A - Viscosity Ranges

Appendix A - Visc

o

sity Ranges

LV and RV,HA,HB Viscometers

Viscosity Range (cP)

Viscometer

Minimum

Maximum

LVDV-E

15

2 M

RVDV-E

100*

13 M

HADV-E

200*

26 M

HBDV-E

800*

106 M

*Minimum viscosity with optional RV/HA/HB-1 spindle

Small Sample Adapter (SSA) and Thermosel (Tsel)

SSA/

Thermosel

Spindle

Shear Rate

(1/SEC)

Viscosity (cP)

LVDV-E

SC4-16 (SSA)

0.29 N

120

-

400 K

SC4-18 (SSA/Tsel)

1.32 N

3

-

10 K

SC4-25 (SSA)

0.22 N

800

-

1.60 K

SC4-31 (SSA/Tsel)

0.34 N

30

-

100 K

SC4-34 (SSA/Tsel)

0.28 N

60

-

200 K

HT-81 (Tsel)

1.29 N

3.5

-

10 K

SC4-82 (SSA)

1.29 N

3.5

-

10 K

SC4-83 (SSA)

1.29 N

11.0

-

38 K

SSA/

Thermosel Spindle

Shear

Rate

(1/SEC)

Viscosity (cP)

RVDV-E

HADV-E

HBDV-E

SC4-14 (SSA)

0.40N

1.25K -

4.2 M

2.5 K -

8.3 M

10 K - 33.3 M

SC4-15 (SSA)

0.48N

500 -

1.7 M

1 K -

3.3 M

4 K - 13.3 M

SC4-21 (SSA/Tsel)

0.93N

50 -

170 K

100 -

300 K

400 -

1.3 M

SC4-27 (SSA/Tsel)

0.34N

250 -

830 K

500 -

1.7 M

2 K -

6.7 M

SC4-28 (SSA/Tsel)

0.28N

500 -

1.7 M

1 K -

3.3 M

4 K - 13.3 M

SC4-29 (SSA/Tsel)

0.25N

1 K -

3.3 M

2 K -

6.7 M

8 K - 26.7 M

HT-81 (Tsel)

1.29N

36 -

10 K

73 -

10 K

292 -

10 K

SC4-82 (SSA)

1.29N

36 -

10 K

73 -

10 K

292 -

10 K

SC4-83 (SSA)

1.29N

121 -

50 K

242 -

50 K

970 -

50 K

cP = Centipoise K = 1,000 M = 1,000,000 N = RPM