Instrument setup – equipment, Operator’s manual – ANKOM RF User Manual
Page 11

Operator’s Manual
Rev F 8/29/14
pg. 11
Instrument Setup – Equipment
Battery Connection
Before using your system you must connect the battery packs within each Module by connecting the male
connector on the battery pack to the female connector on the circuit board. The locking tab must be facing the
adjacent white connector.
Caution: Damage to the circuit board and batteries can occur if the batteries
are connected backwards.
Rechargeable batteries lose charge when not being used.
Always charge your Rechargeable Battery Packs (part # RF16) for at least 3
hours just prior to their use. A fully charged battery when plugged into a
Module should read 6.7 volts or higher on the GPM software screen.
Typically, a fully charged Rechargeable Battery Pack should be able to power
a Module for at least 10 days under normal conditions (20-40°C, GPM
settings of 1 minute Live Interval and 1 Valve Open sequence per hour).
Individual performance may vary depending on testing conditions. Although
Modules will function properly, battery packs may have to be recharged at
more frequent intervals when operating at lower temperatures. When
running the system, remember to check the battery voltage daily and change
the battery pack when the voltage decreases to 6.3 volts or lower. Changing
the battery pack during the run will not affect the results.
Temperature Control
The Module Assembly (includes the Module connected to a Glass Bottle) can be placed in a cabinet incubator or in
a shallow water bath to maintain appropriate temperatures.
Do not allow water into the Module: When placing a Module Assembly in
water, do NOT allow the water level to reach the vent tube or water damage
will occur. When using a water bath, do NOT cover the bath and the RF
modules with a lid as the trapped humidity will be detrimental to the
electronics. Instead, use bath balls to cover the surface of the water and
control the temperature and evaporation.
Battery Connection