Figure 1. Firmware Upgrade Report Sample
•
---------------------------------------------------------------------------
•
UPGRADE FIRMWARE
•
Serial Number: 20050800001 / Model Number: 45NT0
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------
•
Calibration Settings Prior To Firmware Upgrade
•
---------------------------------------------------------------------------
•
Serial Number: 20050800001
•
Model Number: 45NT0
•
Software Versions:
•
Unit : 5.7
•
BP Module: 1.02
•
Temperature Module: 1.2
•
SpO2 Module: 1.9.0.1
•
Pressure:
•
Zero: -96
•
Gain: 18871
•
Cal Signature: ATE2
•
Cal Date/Time: 11/22/2005 07:57:24 PM
•
Voltage:
•
Cal Factor: 16608
•
Cal Signature: ATE2
•
Cal Date/Time: 11/22/2005 07:56:43 PM
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------
•
Filename: C:\SOS\Spot LXi Repair Tools\Applications\ultra_prod_5_7.mot
•
Highest address loaded 0xA3297
•
Erasing Sector 4
•
Erasing Sector 5
•
Erasing Sector 6
•
Erasing Sector 7
•
Erasing Sector 8
•
Erasing Sector 9
•
Erasing Sector 10
•
Erasing Sector 11
•
Erasing Sector 12
•
Erasing Sector 13
•
PROM CRCs sent: 4278684E D738B593
•
Received CRCs 4278684E 4278684E D738B593 D738B593
•
CRC checksums good!
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------
•
Calibration Settings After Firmware Upgrade
•
---------------------------------------------------------------------------
•
Serial Number: 20050800001
•
Model Number: 45NT0
•
Software Versions:
•
Unit : 5.7
•
BP Module: 1.02
•
Temperature Module: 1.2
•
SpO2 Module: 1.9.0.1
•
Pressure:
•
Zero: -96
•
Gain: 18871
•
Cal Signature: ATE2
•
Cal Date/Time: 11/22/2005 07:57:24 PM
•
Voltage:
•
Cal Factor: 16608
•
Cal Signature: ATE2
•
Cal Date/Time: 11/22/2005 07:56:43 PM
•
The above checks are designed to:
•
1.) Verify that the device stored data was not lost.
•
2.) Verify Firmware installation was completed and accurate during the upgrade process.
•
The Device should have Calibration and Operational performance verified prior to first use.
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------
•
Signature Block
•
Welch Allyn Representative __________________ __________________ ______
•
Printed Name Signature
Date
•
Customer Representative __________________ __________________ ______
•
Printed Name Signature
Date
•
---------------------------------------------------------------------------
•
---------------------------------------------------------------------------