Figure 1. firmware upgrade report sample – Welch Allyn Spot LXI Repair Tool Firmware Upgrade Procedure - Installation Guide User Manual

Page 2

background image

Figure 1. Firmware Upgrade Report Sample

---------------------------------------------------------------------------

UPGRADE FIRMWARE

Serial Number: 20050800001 / Model Number: 45NT0

---------------------------------------------------------------------------

---------------------------------------------------------------------------

---------------------------------------------------------------------------

Calibration Settings Prior To Firmware Upgrade

---------------------------------------------------------------------------

Serial Number: 20050800001

Model Number: 45NT0

Software Versions:

Unit : 5.7

BP Module: 1.02

Temperature Module: 1.2

SpO2 Module: 1.9.0.1

Pressure:

Zero: -96

Gain: 18871

Cal Signature: ATE2

Cal Date/Time: 11/22/2005 07:57:24 PM

Voltage:

Cal Factor: 16608

Cal Signature: ATE2

Cal Date/Time: 11/22/2005 07:56:43 PM

---------------------------------------------------------------------------

---------------------------------------------------------------------------

Filename: C:\SOS\Spot LXi Repair Tools\Applications\ultra_prod_5_7.mot

Highest address loaded 0xA3297

Erasing Sector 4

Erasing Sector 5

Erasing Sector 6

Erasing Sector 7

Erasing Sector 8

Erasing Sector 9

Erasing Sector 10

Erasing Sector 11

Erasing Sector 12

Erasing Sector 13

PROM CRCs sent: 4278684E D738B593

Received CRCs 4278684E 4278684E D738B593 D738B593

CRC checksums good!

---------------------------------------------------------------------------

---------------------------------------------------------------------------

Calibration Settings After Firmware Upgrade

---------------------------------------------------------------------------

Serial Number: 20050800001

Model Number: 45NT0

Software Versions:

Unit : 5.7

BP Module: 1.02

Temperature Module: 1.2

SpO2 Module: 1.9.0.1

Pressure:

Zero: -96

Gain: 18871

Cal Signature: ATE2

Cal Date/Time: 11/22/2005 07:57:24 PM

Voltage:

Cal Factor: 16608

Cal Signature: ATE2

Cal Date/Time: 11/22/2005 07:56:43 PM

The above checks are designed to:

1.) Verify that the device stored data was not lost.

2.) Verify Firmware installation was completed and accurate during the upgrade process.

The Device should have Calibration and Operational performance verified prior to first use.

---------------------------------------------------------------------------

---------------------------------------------------------------------------

Signature Block

Welch Allyn Representative __________________ __________________ ______

Printed Name Signature

Date

Customer Representative __________________ __________________ ______

Printed Name Signature

Date

---------------------------------------------------------------------------

---------------------------------------------------------------------------

6.0

C:\Program Files\Welch Allyn\Spot LXi Repair Software\ultra_prod_6_0.mot