WARNING:
It is the health facility or other user’s responsibility to assure that the ULTIMOUNT™ wall mounting
system is anchored to a wall securely and consideration must be given to the load that will be placed
upon the wall mount. Each ULTIMOUNT™ is labeled with a maximum load rating of 33 pounds. Do
not exceed this load limit.
WARNING:
Welch Allyn, Inc. does not accept responsibility for any installation, nor for damage or injury arising from the
installation in any type of wall regardless of the type of fasteners and technique used.
WARNING:
It is the responsibility of the user of the ULTIMOUNT™ wall mount to see that it is securely anchored to the
wall and that the Atlas™ monitor is securely fastened to the wall mount as directed below.
Tools Required:
Phillips head Screw Driver
5/32” Hexagon Wrench (provided)
Hardware Included:
(1) Mounting Plate
(2) #8-32 Pan Head Phillips Screw - 3/8” long
(2) #8-32 Internal Tooth Lockwasher
(4) #10-32 Pan Head Phillips Screw - 1/2” long
(4) #10-32 Internal Tooth Lockwasher
Part No. 620426 Rev. A
Atlas
TM
Wall Mount Installation Guide
Using the ULTIMOUNT™ Wall Mounting System
Mounting Plate
Arm Assembly
Track
Basket
Locking Pin
Locking Pin
Track Stop Screws
Track
Hardware
Locking Pin Holes