Understanding Battery Performance Capabilities
To maximize ProBP battery
performance ‐ and achieve 100
readings per charge with a new
battery ‐ the user must turn power
off after each reading has been
documented.
If power is left on, the device will power
down in 30 minutes*, however
maximum cycles per charge will be
significantly reduced.
Chart based on Average Use Case as described in table below
Battery Charge Cycles
~500
~ 100
Cy
cle
Capacity
(BP
Re
ad
in
gs
)
of
Fully
Char
ge
d
B
atte
ry
†
FACTOR
Decreases Cycle Capacity
of Fully Charged Battery
AVERAGE USE
CASE
Increases Cycle Capacity
of Fully Charged Battery
MODE
N/A
STEP DEFLATION
SureBP Mode
INITIAL INFLATION
PRESSURE
(Step Deflation Mode Only)
Higher initial inflation pressure
160mmHg
Lower initial inflation pressure
CUFF SIZE
Larger blood pressure cuff
ADULT
Smaller blood pressure cuff
BLOOD PRESSURE
Higher Blood Pressure
120/80
Lower blood pressure
BATTERY AGE
Battery retains approximately
75% of original capacity after 500
charge cycles
NEW
N/A
† To fully charge ProBP battery,
continue charging with power
off for 1 hour after charge
status indicator changes from
amber to green.
*
Default settings:
2 Min. – Screen Blank
30 Min. – Power Off
Adjustable using
Welch Allyn
Service Tool
To turn power off after
Screen Blank, press any
button to reactivate display
then press Power Button