Endotracheal Tube Kinked Or Airway Obstructed

If exhaled CO

2

gas is completely or partially obstructed, the slope of the

ascending limb of the capnogram has little or no alveolar plateau. Look for:

Partially kinked or occluded artificial airway

Herniated endotracheal or tracheal tube cuff

Bronchospasm, chronic obstructive pulmonary disease

Foreign body in the upper airway

60

30

0

Kinking

QG_CO2.frm Page 6 Friday, February 13, 2004 4:26 PM