SpO2 Nellcor sensor error codes
Nellcor Errors (Section 3.1.1 of Nell-1 Host Controller Interface Specification [6] - see references)
Recovery
number
Description
Recovery
0x01
Irrecoverable malfunction.
Replace Nell-1 board.
0x02
Irrecoverable malfunction.
Replace Nell-1 board.
0x23
Software error; reset in progress.
Replace Nell-1 board if no communication for 15
seconds.
0x45
Communication error
Contact application developer.
0x67
Defective/bad sensor
Fix sensor.
0xDC
Communication error with sensor
Replace sensor.
0xAB
Recoverable error that needed to be
logged.
Not required.
Action taken for these errors are tabulated in the table below:
Platform SpO2 WUI software errors and recovery
Nellcor recovery
number
Masimo board
failure code
Software generated
error
Host controller
recovery action
0x01
0x02
All
BOARD_FAIL
Display a message to
replace the SpO2 board.
0x23
N/A
BOARD_FAIL_STANDBY Wait 15 seconds for the
communication to restart.
If not, display error to
replace the board.
0x67
0xDC
Upon detection of a
malfunctioning probe.
BAD_PROBE
Display a message
indicating that a
malfunctioning probe is
attached.
0x45
0xAB
N/A
None
N/A
N/A
N/A
WRITE_CONFIG
Check CSpO2CData object
used in the
WRITE_CONFIG command
(see explanation below).
Some configuration may be set by sending the WRITE_CONFIG command. However, if the
configuration specified receives an ACK message, but it fails to apply, the software reports a
WRITE_CONFIG error. This is necessary because it may take up to 2 seconds for the configuration
setting to apply and the module should continue to communicate with the host during this time.
Therefore, the WRITE_CONFIG command, if replied with ACK, fails to apply, a WRITE_CONFIG error
is reported.