beautypg.com

Operating instructions, Maintenance – Welch Allyn Desk Set - Quick Reference Guide User Manual

Page 2

background image

15mfd

High Impedance Voltmeter

1000

15mfd

High Impedance Voltmeter

1000

OPERATING INSTRUCTIONS

SYMBOLS

By plugging into any 220-250 v., 50-60 Hz. AC outlet, the desk
charger is immediately ready for use. An indicator light shows
that the charger is plugged in and ready to charge either one
or two Numbers 60713, 71500, or 71670 battery handles.

By leaving the handles in the charger, fully-charged instruments
are always ready for use. Batteries cannot be over-charged.
A completely discharged battery handle can be fully charged
overnight.

Attention: Consult Manual

The mark on this device indicates that it has been tested to and conforms with the provisions
noted within the 93/42/EEC Medical Device Directive.

Authorized European Representative

European Regulatory Manager
Welch Allyn, LTD
Kells Road, Navan
County Mead
Republic of Ireland
Tel: 353 46 28122
Fax: 353 46 28536

CAUTION: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.

MAINTENANCE

LEAKAGE CURRENT CHECK

EARTH CONTINUITY CHECK

CLEANING

Like any other piece of electrical equipment, periodic electrical inspections should be
performed by qualified personnel. We recommend that this equipment be inspected at
least every six months, more often if used under adverse conditions.

The leakage current from exposed metal parts, when a good earth connection is used, should be
less than 10 microamps. This equivalent to 9.5 mv in the following circuit.

The resistance measured from the earth pin of the plug of the bottom plate should be virtually zero.

Do

try to sterilize. May be cleaned by wiping with a dry cloth. Do

pour any liquids into the

charging wells.

For repairs by qualified personnel, contact your local Welch Allyn distributor.

not

not

OPERATING INSTRUCTIONS

SYMBOLS

By plugging into any 220-250 v., 50-60 Hz. AC outlet, the desk
charger is immediately ready for use. An indicator light shows
that the charger is plugged in and ready to charge either one
or two Numbers 60713, 71500, or 71670 battery handles.

By leaving the handles in the charger, fully-charged instruments
are always ready for use. Batteries cannot be over-charged.
A completely discharged battery handle can be fully charged
overnight.

Attention: Consult Manual

The mark on this device indicates that it has been tested to and conforms with the provisions
noted within the 93/42/EEC Medical Device Directive.

Authorized European Representative

European Regulatory Manager
Welch Allyn, LTD
Kells Road, Navan
County Mead
Republic of Ireland
Tel: 353 46 28122
Fax: 353 46 28536

CAUTION: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.

MAINTENANCE

LEAKAGE CURRENT CHECK

EARTH CONTINUITY CHECK

CLEANING

Like any other piece of electrical equipment, periodic electrical inspections should be
performed by qualified personnel. We recommend that this equipment be inspected at
least every six months, more often if used under adverse conditions.

The leakage current from exposed metal parts, when a good earth connection is used, should be
less than 10 microamps. This equivalent to 9.5 mv in the following circuit.

The resistance measured from the earth pin of the plug of the bottom plate should be virtually zero.

Do

try to sterilize. May be cleaned by wiping with a dry cloth. Do

pour any liquids into the

charging wells.

For repairs by qualified personnel, contact your local Welch Allyn distributor.

not

not