Welch Allyn Autorefractor Quick Reference Guide - Quick Reference Guide User Manual
Quick reference guide, Pre-test set-up, Test procedure

Quick Reference Guide
Pre-test Set-up:
•
Push any button to turn the unit on.
•
Position the patient so that the test can be conducted level with, and square to, patient’s eyes.
•
Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double
beep and see the desired icon on the LCD.
•
Explain the test procedure to the patient:
•
“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”
•
Position yourself at eye-level and square with the face of the patient.
Test Procedure:
•
Push the “GO” button on the unit
•
Check that the patient is fixated correctly throughout the test.
•
Position the unit at the correct distance from the patient:
•
When the unit is too far away, you will hear slow, low-pitched beeps.
• • • • • •
•
Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———
•
When the unit is too close, you will hear quick, high-pitched beeps.
••••••
•
The crosshair will flash in synch with the tones.
•
Look through the peephole and align the crosshair on the pupil of the patient’s right eye. While the unit is acquiring data, you
will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °
•
If you are not acquiring data (steady tone without acquiring data chirps):
•
Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.
•
Ask the patient if they can see the red light.
•
When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not
need to press any buttons for the left eye test.)
•
Without changing position, rotate the unit to the left eye and align the crosshair over the pupil. Repeat test.
•
At the end of the test, you will hear the “tah-dah” sound again.
•
If the unit has not gathered enough good readings from either eye, you
will hear 5 tones when the test stops. - - - - - You must re-test that eye.
•
To stop a test at any time, hold any button until 5 tones sound.
Hints To Insure Good Results & Rapid Test:
•
Ambient lighting:
•
Don’t perform by uncovered windows
•
Dimmed light can help for those with small pupils
•
Too dim is difficult for those with dark irises
•
Entry angle - straight and level
•
Rotate unit to test left eye
•
Fixation: “Look at the red light”
•
Remind and monitor
•
Match patient age and mode
•
These are noted on the print out
•
Correct distance/Not taking readings (steady tone without chirps):
•
Ensure you are level and square with the patient
•
Scan the crosshair around the eye in an outward moving spiral until
chirps begin, hold this location
•
Ask patient if they can see the red light
•
Aim high if testing through glasses
•
Make sure the patient’s eyelids are not occluding the pupil
Quick Reference Guide
Pre-test Set-up:
•
Push any button to turn the unit on.
•
Position the patient so that the test can be conducted level with, and square to, patient’s eyes.
•
Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double
beep and see the desired icon on the LCD.
•
Explain the test procedure to the patient:
•
“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”
•
Position yourself at eye-level and square with the face of the patient.
Test Procedure:
•
Push the “GO” button on the unit
•
Check that the patient is fixated correctly throughout the test.
•
Position the unit at the correct distance from the patient:
•
When the unit is too far away, you will hear slow, low-pitched beeps.
• • • • • •
•
Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———
•
When the unit is too close, you will hear quick, high-pitched beeps.
••••••
•
The crosshair will flash in synch with the tones.
•
Look through the peephole and align the crosshair on the pupil of the patient’s right eye. While the unit is acquiring data, you
will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °
•
If you are not acquiring data (steady tone without acquiring data chirps):
•
Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.
•
Ask the patient if they can see the red light.
•
When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not
need to press any buttons for the left eye test.)
•
Without changing position, rotate the unit to the left eye and align the crosshair over the pupil. Repeat test.
•
At the end of the test, you will hear the “tah-dah” sound again.
•
If the unit has not gathered enough good readings from either eye, you
will hear 5 tones when the test stops. - - - - - You must re-test that eye.
•
To stop a test at any time, hold any button until 5 tones sound.
Hints To Insure Good Results & Rapid Test:
•
Ambient lighting:
•
Don’t perform by uncovered windows
•
Dimmed light can help for those with small pupils
•
Too dim is difficult for those with dark irises
•
Entry angle - straight and level
•
Rotate unit to test left eye
•
Fixation: “Look at the red light”
•
Remind and monitor
•
Match patient age and mode
•
These are noted on the print out
•
Correct distance/Not taking readings (steady tone without chirps):
•
Ensure you are level and square with the patient
•
Scan the crosshair around the eye in an outward moving spiral until
chirps begin, hold this location
•
Ask patient if they can see the red light
•
Aim high if testing through glasses
•
Make sure the patient’s eyelids are not occluding the pupil
TM
R
Left/Right Toggle & Print
• Selects eye to re-test
• Prints results (hold button down
while aiming patient side of
SureSight™ at printer until you
hear “tah-dah”)
Clear and Child/Adult
Calibration
• Clears readings
• Changes child/adult
(hold down)
“GO” Button
• Unit on: starts test
• Unit off: recalls old
readings
TM
R
Left/Right Toggle & Print
• Selects eye to re-test
• Prints results (hold button down
while aiming patient side of
SureSight™ at printer until you
hear “tah-dah”)
Clear and Child/Adult
Calibration
• Clears readings
• Changes child/adult
(hold down)
“GO” Button
• Unit on: starts test
• Unit off: recalls old
readings