Welch Allyn Spot Vital Signs LXI - Quick Reference Guide User Manual
Spot vital signs lxi operating instructions
Table of contents
Document Outline
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Spot Vital Signs LXi Operating Instructions
- Blood Pressure Operation
- SureTemp Plus Temperature Operation
- 1. Remove probe from well and apply a new probe cover. Use the Navigation buttons to change the temperature mode then press the Select button.
- 2. Place probe in contact with patient tissue as directed in the Directions for Use manual. When done, a beep sounds. The temperature appears on the display.
- 3. Eject probe cover. Return probe to well before starting another measurement.
- Braun ThermoScan® PRO 4000 Temperature Operation
- 1. Remove probe from holder and apply a new probe cover; when ready, the device will beep.
- 2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
- 3. Press the Activation button. When done, a long beep sounds. The probe will display the temperature.
- 4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will display the temperature.
- SpO2 Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation
- Save/Send Operation
- 1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window.
- 2. Press the Select button. Spot Vital Signs LXi will transfer data electronically if it can or will save data in memory if it cannot transfer it. A blank status window will appear.
- Manual Weight, Height, Respiration, Pain Entry
- Memory/Print/Erase Operation