Welch Allyn MicroTymp3 - Quick Reference Guide User Manual

Interpreting 226-hz results, Interpreting 1000-hz results, Test instructions for the welch allyn microtymp

background image

Outer ear

External ear canal

Semicircular canals

Epitympanic recess

Incus

Malleus

Tympanic membrane

Tympanic cavity

Vestibule

Endolymphatic
duct and sac

Cochlea

Normal middle ear

Positive middle ear

pressure

• Acute otitis media

PASS–Peak Ya

0.6 mmho

Low admittance–Normal

peak Ya, abnormal gradient

• Otitis media with effusion
• Tympanosclerosis

Negative middle ear

pressure

• Eustachian tube dysfunction
• Cold, allergies or vigorous sniffing

REFER–Peak Ya < 0.6 mmho

• Middle ear effusion

Low admittance–Low peak Ya

• Otitis media with effusion
• Tympanosclerosis
• Cholesteatoma
• Middle ear tumor

Flat tympanogram–

Normal ear canal volume

• Middle ear effusion

RETEST–Try again!

High admittance–High peak Ya

• Tympanic membrane

abnormalities

• Ossicular disruption

Flat tympanogram–

Increased ear canal volume

• Patent tympanostomy tube
• Perforated tympanic membrane

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

RIGHT EAR

R

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

RIGHT EAR

R

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

LEFT EAR

L

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

LEFT EAR

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

LEFT EAR

L

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

LEFT EAR

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

RIGHT EAR

R

High Peak Ya

-400

-200

0

200

2

0.5

0

1.0

1

1.5

PRESSURE - daPa

Ya - mmho

Vea

+200

LEFT EAR

NOTE: Multiple RETEST results

can be considered a REFER.

InTErPrETIng 226-Hz rEsulTs

(tympanograms and associated clinical conditions)

InTErPrETIng 1000-Hz rEsulTs

Stapes

Eustachian tube

Seal

Tip

Tip ejector

Probe

TIPs fOr suCCEss

BIggER IS BETTER

When using a tympanometric device with a probe designed

for an external seal, such as the MicroTymp3, a bigger tip is

generally better since the seal is obtained from outside the
external auditory canal, not inside.

USE THE “ScOOP” METHOd

Attain best result by “scooping” the

probe in behind the tragus.

TRY A FAMILIAR ANgLE

The head of the MicroTymp3 handle is

angled at 15°— the same angle as a

Welch Allyn otoscope. Try holding

the MicroTymp3 at the same angle
required for an otoscopic view.

PRAcTIcE,

PRAcTIcE, PRAcTIcE

For those who are new to

tympanometry, or to this type of

hand-held device, there are benefits to

practicing the technique a few times. This builds

confidence and skill, and makes getting a seal quite easy.

TEsT InsTruCTIOns fOr THE WElCH AllYn MICrOTYMP

®

3

SELEcT A TIP

1. Select a large enough tip to seal the entrance to the ear canal.
2. Push the tip onto the probe. Make sure it is fully seated.
TEST

3. Pick up the handle and press . Either “OPEN 226” or “OPEN 1000” will display.
4. To switch to the other frequency, press both the Right memory and

Left memory buttons on the handle simultaneously until the desired frequency

displays. Press to confirm the frequency.

5. Grasp the patient’s pinna. Pull gently back to straighten the ear canal. Both you

and the patient should be motionless during the test.

6. Press the tip firmly against the ear canal opening. Once you achieve a seal, the

TEST message displays. The test is complete when the last data point (226 Hz)

or when a “Quick-Interpretation” message (1000 Hz) displays.

STORE/REcORd

7. 226 Hz: Store test data in the handle by pressing the memory button for the tested

ear: or .

1000 Hz: Record the “Quick-Interpretation” message and value on the MicroTymp3

Screening Results card.

PRINT (226 Hz)

8. Place the handle in the Printer/Charger. The green CHARGE indicator illuminates.
9. Press .

TEST

TEST

R

R

L

L