English, Set the date and time

1
English
Set the date and time
After recharging a dead battery or after disconnecting the battery for a few minutes, you
must program the date and time screen.
You can change several device operating parameters in the Internal Configuration Mode.
When changed, these settings become the default power-up settings. You will also see
non-changeable device configurations for technical service purposes.
To enter the Internal Configuration Mode:
1.
Turn the Spot Vital Signs off (
).
2.
Press and hold the Blood Pressure Start/Stop (
) and Power (
) buttons. The
device enters the Internal Configuration Mode and displays the software version.
If the device displays E38, press the Next Patient/Clear/Cancel (
) button to
cancel the error.
3.
Press the Mode (
) button to cycle through the Internal Configuration menu until
you see the date option displayed on the screen.
4.
Use the Next Patient/Clear/Cancel (
) or Blood Pressure Start/Stop (
)
buttons to change the default setting.