Charging instructions, Low battery indicator, Battery replacement – Welch Allyn 719 Series Lithium Ion Handle - User Manual User Manual
Page 2: Cleaning and disinfection, Specifications, Service and warranty, Symbols
![background image](/manuals/641778/2/background.png)
CHARGING INSTRUCTIONS
1.
Remove instrument head.
2.
Insert T-stud on the AC Charging Module into the T-slot on the Battery Section (Figure 2). Rotate
90° in either direction. T-stud will lock into place.
3.
To expose prongs, depress lock Button on the AC Charging Module (Figure 1). With the Lock
Button depressed, slide Sleeve toward Rheostat Section.
4.
Plug into standard (110-120v 50/60 Hz) wall outlet. Charge for 18 continuous hours to charge a
dead battery.
5.
Battery may be charged after partial discharge without negatively affecting overall battery
memory (unlike Ni-Cad cells).
6.
Do not attempt to use the instrument during charge, the instrument will not operate. When
charging is complete remove the handle and charger from outlet. For longest service, do not
leave AC Charging Module plugged into a power outlet without a handle attached.
7.
Slide sleeve back over prongs. The Lock Button will click when prongs are fully hidden.
8.
Remove AC Charging Module from the Battery Section if desired.
CAUTION The battery section is not shipped fully charged. For maximum performance charge the
battery section for 12 - 16 hours when battery section is new or after a long period of inactivity.
LOW BATTERY INDICATOR
1.
The Low Battery Indicator will illuminate several minutes before handle shuts off depending on
the age of the battery (Figure 1).
2.
Once the battery is fully discharged, the Low Battery Indicator may flicker.
BATTERY REPLACEMENT
Removing the Battery Section:
1.
Hold the instrument handle firmly by the Battery Section
2.
Turn the Rheostat Section counterclockwise until it detaches from the Battery Section.
Replacing a New Battery Section
1.
Insert the Rheostat Section into the new Battery Section.
2.
To connect the two sections, turn the Rheostat Section clockwise until tight.
3.
Verify the Rheostat Section is secure on the Battery Section of the instrument handle.
CLEANING AND DISINFECTION
Wipe the handle with an appropriate health care low- or intermediate-level cleaner/disinfecting wipe
that incorporates either a 1:10 sodium hypochlorite (bleach) solution or isopropyl alcohol as the active
disinfection ingredient. Follow wipe manufacturer’s instructions for appropriate use, contact times and
applicable warnings and precautions.
Do not excessively soak/saturate the handle assembly.
Do not immerse handle assembly in any solution.
Do not sterilize the Power Handle.
After disinfection, inspect the handle assembly for visible signs of deterioration at its switch, device
end connector and housing. If evidence of damage or deterioration is present, discontinue use and
contact Welch Allyn or your area representative.
SPECIFICATIONS
(Specifications and design may change without notice)
Model Number:
71900
Dimensions:
7.5” Long x 1.125” Diameter (Max.), 5.4” Long without AC Charging Module
Weight:
7.1 oz, 4.5 oz without AC Charging Module
Typical Charge Time: 18 hours (fully discharged)
Typical Discharge Time:110 minutes (for new battery)
Input:
120V~60Hz 0.17A
Conforms with:
IEC/UL/CSA/EN 60601-1
Continuous Operation
SYMBOLS
SERVICE AND WARRANTY
Service
If the product fails to function properly or if assistance, service, or spare parts are required, contact the
nearest Welch Allyn Technical Support Center.
Welch Allyn Technical Support:
http://www.welchallyn.com/about/company/locations.htm
Warranty
The 719 series Power Handle is guaranteed by Welch Allyn against all manufacturing defects. Welch
Allyn will repair or replace, free of charge, any parts of its own manufacture proven to be defective
through causes other than misuse, neglect, damage in shipment, or normal wear.
Welch Allyn warrants that the Welch Allyn 71960 Battery Section will perform to original specifications
during the two years from the date of manufacture, when used with the Welch Allyn 719 series Power
Handle. This warranty does not apply to the use of the 71960 battery in other products. A defective
battery will be replaced if it fails within two years of the date of manufacture.
Battery Recycling
Welch Allyn now employs the services of an agency which can disassemble and recycle all components
of rechargeable batteries so that nothing gets landfilled or incinerated. Should you wish to dispose of
your expended Welch Allyn rechargeable battery via recycling (in the U.S. and Canada only), please
send to:
Welch Allyn
Welch Allyn Canada
RECYCLE BATTERY
RECYCLE BATTERY
4341 State Street Road
160 Matheson Blvd. East
Skaneateles Falls, NY
Mississauga, Ontario
13153-0220
CANADA L4Z 1V4
For locations outside the U.S. and Canada consult your local recycling authority for correct disposal of
batteries.
Temperature limitation
Operating:
59°F (10°C) -104°F (49°C)
Transport/storage:
-4°F (-20°C ) - 120°F (55°C)
Humidity limitation
Operating: 30% - 90%
Transport/storage: 10% - 95%
Atmospheric pressure limitation
500 hPa - 1060 hPa
Attention. Read Operating Manual for Cautions and Instructions for Use.
Type BF applied part
Class II equipment
Reorder number
Product Identifier
Manufacturer & Date of Manufacture (YYYY-MM-DD)
Professional use only
Global Trade Item Number
Consult operating instructions
Batch code
Equipment not protected against the ingress of water.