beautypg.com

Product specifications – Welch Allyn PRO 4000 Base Station - User Manual User Manual

Page 8

background image

8

• When you release the start-button, the thermo-

meter‘s display will show the actual anti-theft time
setting. E.g. «H08» meaning 8 hours have been
selected as the time for which the thermometer
can be used before it has to be returned to the
station.

• To change the anti-theft interval setting, press the

start button again and as often as necessary until
the desired time is shown in the display.

«H--» means the anti-theft function is off, «H01»
means 1 hour setting, etc.

• Release the start button and place the thermo-

meter back into the station within 30 seconds.
A beep will sound, confirming the setting.

• If the setting is not confirmed, repeat the proce-

dure as described.

If the anti-theft function is enabled, and when not
in the measuring process (stand-by mode) the
thermometer will show the time remaining before
the anti-theft function shuts down its operation.
«H3.5» means the anti-theft function will shut down in
3.5 hours.

If the thermometer shows «RET» (without the battery
symbol) on its display, the anti-theft time has elapsed
and the thermometer must be returned to the station
for unlocking.
If the anti-theft function is not enabled, the display
will show «OFF» when in stand-by mode.

Product specifications

Long term storage ranges

Temperature:

–4 °F to 122 °F
(–20 °C to 50 °C)

Humidity:

95% non condensing

Shock:

withstands drop of 3 feet

Rechargeable Battery life: 3 years, 1000 charging

cycles

Battery capacity:

1 month standby,
500 measuring cycles

Rechargeable Battery type: customized dual battery

pack (NiMH)

6021361_Pro4000_Base Seite 8 Freitag, 6. Juli 2007 10:11 10