Trouble Shooting Guide

Problem

Possible Cause

Solution

Lamp does not
light.

Rectal Light Handle is not
plugged into the power
supply.

Verify that the Rectal Light Handle is
properly plugged into power supply
cord.

Power Supply is not plugged
in.

Verify the power supply is properly
plugged into the outlet.

Faulty Lamp

Replace with a new lamp.

Loose Lamp

Verify that the lamp is secured into the
Rectal Light Handle.

Wrong Lamp

Verify that the lamp is a REF 07800
Welch Allyn lamp.

Damaged Lamp

Inspect the lamp and the o-ring for
damage. Replace with REF 07800
lamp if damage is found.

Rectal Light
Handle does not
stay plugged into
the transformer
cord.

Rectal Light Handle/
transformer interconnect
plug is damaged.

Inspect interconnect plug for damage.
Replace either or both REF 73210 or
REF 73305, 73322, 73324, 73326.

Light output
seems low.

Faulty Lamp

Replace with a new lamp.

Dirty lamp lens

Inspect lamp lens for debris and clean.

This guide is intended to provide assistance towards determining the fault condition of
this product. Please contact the authorized Welch Allyn Service Center listed within the
product owners manual for additional assistance.