990070 Rev. C

Specifications

Power Requirements:

100-127 volt,

~

, 60 Hz., 1.75A

220-240 volt,

~

, 50 Hz., 650 mA

IEC 320 socket with removable cord is located at back; verify unit has correct

voltage before use

Air Flow:

80 cfm (low) or 120 cfm (high)
(2.26 or 3.39 m

3

/min)

Approximate Size: 13” x 13” x 27” (33 x 33 x 68.6 cm)

Approximate Weight:

45 lbs. (20.4 kg)

Control Panel:

located on door of instrument for easy viewing

Filters:

HEPA DOP 99.97% / 99.99% and activated carbon
pre-filter

Filter Replacement: user access through door

Hose:

3” dia. (7.6 cm), 3’ (91 cm). length with expanded

intake

Power Switch:

located adjacent to power inlet

Environment Limitations:

Operational Requirements:

Temperature: 50

°

to 95

°

F (+10

°

to 35

°

C)

Humidity: 30 % to 75 %
Pressure range: 10.15 to15.37 psia (70 to 106 kPa)

Transport and Storage Requirements:

Temperature: 14

°

to 122

°

F (-10

°

to 50

°

C)

Humidity: 10 % to 90 %
Pressure range: 7.25 to 15.37 psia (50 to 106 kPa)

Equipment Classification:

the Kreativ Model Klean Air II system is an EC
Class I device requiring an electrical earth ground.
This device is classified as ordinary with regard
to protection against water ingress.

Electrical Fuses:

Littelfuse, SlowBlow

Model

Fuse Mfr.

Mfr. P/N

Kreativ P/N

Fuse Rating

Klean Air II
Domestic

Littelfuse

Type 218-02.5

600109

2.5A, 250Vac

Klean Air II
European

Littelfuse

Type 218-01.6

600096

1.6A, 250Vac