DESCRIPTION

Vaginal specula are available from W

elch Allyn’

s line that allow use in almost any situation.

There are single-use and re-usable plastic specula for general office or operating room use, as well as Rose

and Prima Series

electrosurgical specula. All of the rose-coated and Prima

specula are non-conductive and

autoclavable. Many of the styles include either permanent or disposable smoke evacuator tubing.

INDICA

TIONS FOR USE

Vaginal specula are indicated whenever a visual examination of the vagina and cervix is

necessary

. The instruments are also necessary for visualization when treating

vaginal/cervical lesions.

S

ome of the indications include (not a complete list):

1

General pelvic exam—visualization of the cervix and vagina, and for obtaining pap smears.

2

V

isualization of cervical or vaginal secretions for obtaining cultures or wet smears.

3

V

isualization for evaluation of the source of bleeding: vaginal, cervical, or uterine.

4

Evaluation of the vagina and cervix in presence of vulvar or perineal viral

(Herpes or HPV) lesions.

5

V

isualization of cervix and or vagina as part of evaluation and treatment of CIN; Abnormal

pap, V

AIN, HPV or other cervical lesions.

6

In conjunction with biopsy

, electrosurgical, cryo or cold knife treatment methods.

7

Follow-up evaluation of treatment of cervical or vaginal lesions.

8

Preoperative evaluation of vaginal and uterine supports relating to descent and prolapse.

9

Post operative evaluation of vaginal apex after abdominal or vaginal hysterectomy

.

10

Post operative evaluation after vaginal or abdominal pelvic relaxation surgical procedures.

CON

TRAINDICA

TIONS

Please refer to hysteroscope or electrosurgical generator manuals.

W
ARNINGS

Inform the patient when the speculum is introduced and removed.

Adjust as necessary to provide improved comfort.

1

6

ETHYLENE O

X

IDE

n

Thorouthly clean specula of all foreign matter prior to sterilization, following steps A and B.

n

Follow the manufacturer’

s instructions for operation and loading of sterilizer

.

During sterilization provide direct exposure to all surfaces of the instruments.

n

Sterilize instruments in their “open” position.

n

A

void contact of plastic to bare metal.

RECOMMENDED HOSPIT

AL EHTYLENE O

XIDE CY

CLE

Temperature

125-130°F

50% RH (pre-humidity)

60 minutes

-0/+10 minutes

Pre V

acuum

24” Hg

±

2” Hg

Gas Pressure

6-8 psig (550-660 mg/L)

Exposure T

ime

4 hours -0/+0.25 hours

Post V

acuum

24” Hg 2X

±

2” Hg

Aeration

12-0/+1 hours at 120°F