Specifications, Classification discarding the equipment – Welch Allyn ABPM 6100 Addendum User Manual
Page 2

2
ABPM 6100 Addendum
Safety and Effectiveness Warnings and Cautions
Make sure that you are familiar with all Safety and Effectiveness considerations before
using the ABPM 6100.
Specifications
This section provides additional specifications for the ABPM 6100.
Classification
Discarding the Equipment
Certified to: CAN/CSA STD C22.2 NO. 601.1
Conforms to: IEC 60601-1, 60601-1-2, UL STD 2601-1
Recycle
WARNING The
ABPM 6100 has not been designed for use with high
frequency (HF) surgical equipment and does not protect against hazards to
the patient.
WARNING Ensure the location of the ABPM 6100 provides maximum
separation away from all sources of high-frequency energy.
Environmental Operating
Condition:
Atmospheric Pressure Range of 700 HPA to 1060 HPA
Protection against electric
shock:
Internally powered, Type BF
Protection against electric
shock:
Internally powered, Type BF
Protection against the
ingress of liquids:
IPXØ, per IEC 60529
Discard the old battery appropriately.
•
In the USA, call 1800-SAV-LEAD for instructions on recycling it.
•
International users, contact your local authorities concerning
recycling.
•
Discard the ABPM 6100 and accessories according to local laws.
Do not dispose of this product as unsorted municipal waste. Prepare this
product for reuse or separate collection as specified by Directive 2002/
96/EC of the European Parliament and the Council of the European Union
on Waste Electronic and Electrical Equipment (WEEE). If this product is
contaminated, this directive does not apply. For more specific disposal
information, see www.welchallyn.com/weee, or contact Welch Allyn
Customer Service.