Material 711686
80014418, Ver. C
IMPORTANT INFORMATION ABOUT YOUR OAE HEARING SCREENER
If you have purchased the Seiko Smart Label Printer Model 440 or 650SE to use with a Welch Allyn OAE Hearing
Screener, you may need to upgrade the firmware within the screener module in order to successfully use this
printer. Please read the Technical notes below to determine whether this applies to your system.
Welch Allyn Load Version 4.4.2.i01 Technical Notes
Release date: December 2013
The change was made to add support for the Seiko Smart Label Printer Model 440 or 650SE for the Welch Allyn
OAE Hearing Screener. To be able to use the Model 440 or 650SE Seiko printer, customers MUST have the
firmware associated with Welch Allyn Load version 4.4.2.i01 installed on the OAE screener module. A CD with
this version of Welch Allyn Load is provided in the printer box for your convenience.
HOW TO DETERMINE WHETHER YOU NEED TO UPGRADE THE FIRMWARE
To understand whether your OAE Hearing Screener has a compatible version of firmware already installed:
1. Turn on your OAE Hearing Screener unit
2. Press the Select key
3. Watch carefully for the display to show a message in the format:
Loading #.##.##
PLEASE WAIT…
4. Note the number displayed in the #.##.## location noted above.
5. If your system displays a number lower than 4.01.00, then you WILL NEED to upgrade the firmware in the
module.
6. If your system displays the number 4.01.00, then you can use only the Model 440 printer without a
firmware upgrade. In this case, you WILL NEED to upgrade the firmware in the module in order to use
Model 650SE label printer.
7. If your system displays the number 4.02.00 or higher, then you can use either of the Model 440 or 650SE
printers without a firmware upgrade.
Turn page over to see instructions about HOW TO UPGRADE THE Welch Allyn OAE Hearing Screener