beautypg.com

Welch Allyn GS 600 Minor Procedure Light - User Manual User Manual

Gs 600 minor procedure light, Symbol descriptions, Warnings and cautions

background image

Symbol descriptions

Warning (the background

See the accompanying

of this symbol is

manual

Yellow on the device)

Caution: High intensity

Caution

light.

Temperature limits

Use indoors.

Power on/off.

Atmospheric
pressure

Full power

limitation

Half power.

Humidity limitation

Warnings and cautions

WARNING: Read and understand the instructions prior to use of
this product. Failure to understand the operating requirements for
this product may result in injury to yourself, the patient and/or may
damage the instrument.

WARNING: Do not operate this product in the presence of
flammable anesthetics or other flammable substances in
combination with air, oxygen-enriched environments, or nitrous
oxide; explosion can result.

WARNING: Illumination is intense. Do not stare at the light source
in operation. May be harmful to eyes. To minimize risk of harm to the
eyes from optical radiation hazards, avoid looking at bright light
sources and their reflections, and protect eyes where normal pupil
sizes and aversion responses are not present.

WARNING: Use the examination light in its intended working range
of 20 in (50 cm). Exposures at closer distances may be harmful to
skin. To minimize risk of harm to the skin from optical radiation
hazards, minimize illumination intensity at the tissue examination
site, minimize exposure times, and take additional precautions when
skin sensitivity has been altered through tissue trauma or the use of
anesthesia.

WARNING: This is a class A product. In a hospital/physician's office
environment this product may cause radio interference in which case
the user may be required to take adequate measures.

WARNING: Do not modify this equipment. Any modification of this
equipment can lead to patient injury. Any modification of this
equipment voids the product warranty.

CAUTION: Only authorized Welch Allyn Service Centers can perform
service on this product.

CAUTION: There are no user serviceable/replaceable parts.

CAUTION: Rx only.

Intended use

The Welch Allyn GS 600 is designed to meet the various needs of the
physician’s office, hospital environment and specialist’s office. It is not
intended for rendering diagnosis or surgery.

Mounting

Table/wall mount

WARNING: Ensure that the mounting bracket is securely anchored
to the wall or table, or that the mobile stand is securely assembled,
and the exam light is securely fastened to the bracket or stand as
directed. Do not hang or place objects on the equipment.

Setup instructions

Mobile stand assembly instructions

Set up

Cleaning

WARNING: Ensure that the device is unplugged from wall socket.

Luminaire

Clean with a damp cloth using warm water and mild detergent or one of the
following solutions:

70 percent isopropyl alcohol

10 percent mild bleach solution

Wipe dry with a clean cloth.

Troubleshooting

Warranty

Three years.

Specifications

Electrical
Input: 100-240 V~ 50Hz-60Hz, 0.65A max

Environment
Operating:

Transport/storage:

Equipment classification
Safety Class I, Continuous Operation

Additional technical information
www.welchallyn.com/gsseries

Approvals
IEC 60601-1, UL 60601-1, IEC 60601-1-2, CAN/CSA C22.2 No 601.1-M90,
AS/NZS3200

Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com

Regulatory Affairs Representative
Welch Allyn Limited
Navan Business Park
Dublin Road
Navan, County Meath Republic of Ireland

NOTE: This product complies with current required standards for
electromagnetic interference and should not present problems to other
equipment or be affected by other devices. As a precaution, avoid using this
device in close proximity to other equipment.

www.welchallyn.com/weee

For patent information, please visit www.welchallyn.com/patents.

GS 600
Minor Procedure
Light

Manual part no. 718917
© 2014 Welch Allyn

50°F
(10°C)

95°F
(35°C)

500 hPa

1060 hPa

90%

15%

1.

2.

3.

2.

1.

4.

3.

Possible cause

Solution

Power supply fault.

No power supplied to

light.

Verify all cables are

connected correctly.

Light is not

functioning correctly

Call Welch Allyn Customer

Service.

Green light not
illuminated

Green light
illuminated

50°F (10°C)

95°F (35°C)

90%

15%

700 hPa

1060 hPa

50°F (10°C)

95°F (35°C)

90%

15%

500 hPa

1060 hPa