Directions for use
Welch Allyn fiber optic laryngoscope blade assemblies
Maintenance instructions
Replace the fiber optic light carrier (REF: 6906X, 6921X, and 6806X)
To replace the fiber optic light carrier:
1. Remove the locking screw by rotating counterclockwise with a standard screwdriver.
2. Pull light carrier away from base of laryngoscope and slide distal end of carrier out of
blade. See Figure 2.
3. Position new carrier and replace locking screw.
4. Rotate locking screw clockwise until secure.
5. Verify lamp and blade engagement/operation using a known working test handle.
6. Reprocess repaired assembly as appropriate per these instructions.
Figure 2
Specifications
Electrical:
For use with ISO 7376-3 (green code) handles only
For information about electromagnetic compatibility (EMC) see Welch Allyn website:
http://www.welchallyn.com
Operating:
Storage/Transport:
32 F
(0 C)
104 F
(40 C)
-4 F
(-20 C)
120 F
(49 C)
Approvals:
Conforms to ASTM F 1195 and ISO-7376-3, IEC/EN 60601-1, IEC/EN 60601-1-2
The CE mark on this product indicates that it has been tested to and conforms
with the provisions noted within the 93/42/EEC Medical Device Directive.
Complies with EMC Framework of Australia
Warranty: One year
Service Information:
For Technical Support or to obtain information about any Welch Allyn product, contact
Welch Allyn Technical Support: www.welchallyn.com/support.
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com
Authorized European Representative Address:
Regulatory Affairs Representative
Welch Allyn, Limited
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland