Model 7815 Parcel Bench Scale User’s Manual
15
5. Press the ZERO key to scroll through the
available capacity selections. When the
choice you want is displayed, press the
TEST key.
6. If LB was selected, the unit will display
CLASS. Press the ZERO key to toggle
between SCALE and CLASS. When the
choice you want is displayed, press the
TEST key. . .
LOAD 0 is displayed.
7. Clear all weight from the weigh platter and
press the TEST key. After a brief wait…
LOAD 150 is displayed (or the current
capacity selected in Step 5). Alternate
calibration points can be chosen using
the ZERO key to scroll choices.
The NCI 7815 allows calibration using less
than full capacity weights. Below are the
alternative weights that can be used to
calibrate your scale for its designed capacity.
Alternative
Capacity
Calibration Weights
150 lb
10 lb, 50 lb
75 kg
10 kg, 50 kg
8. Place chosen calibration weight on the weigh
platter and press the TEST key…
After a brief wait, DONE is displayed.
9. Press SW-1…
The unit returns to normal weighing
mode.
The unit is now tested, configured and calibrated.
It is ready for use in your application.
Make sure the weigh
platter and access
plate are on the scale
to establish proper
zero load.
Table 1
Alternative
Calibration Points
The calibration
weight should be
removed from the
weigh platter before
countdown com-
pletes.