Model 7815 Parcel Bench Scale User’s Manual

15

5. Press the ZERO key to scroll through the

available capacity selections. When the
choice you want is displayed, press the
TEST key.

6. If LB was selected, the unit will display

CLASS. Press the ZERO key to toggle
between SCALE and CLASS. When the
choice you want is displayed, press the
TEST key. . .

LOAD 0 is displayed.

7. Clear all weight from the weigh platter and

press the TEST key. After a brief wait…

LOAD 150 is displayed (or the current
capacity selected in Step 5). Alternate
calibration points can be chosen using
the ZERO key to scroll choices.

The NCI 7815 allows calibration using less
than full capacity weights. Below are the
alternative weights that can be used to
calibrate your scale for its designed capacity.

Alternative

Capacity

Calibration Weights

150 lb

10 lb, 50 lb

75 kg

10 kg, 50 kg

8. Place chosen calibration weight on the weigh

platter and press the TEST key…

After a brief wait, DONE is displayed.

9. Press SW-1…

The unit returns to normal weighing
mode.

The unit is now tested, configured and calibrated.
It is ready for use in your application.

Make sure the weigh
platter and access
plate are on the scale
to establish proper
zero load.

Table 1

Alternative

Calibration Points

The calibration
weight should be
removed from the
weigh platter before
countdown com-
pletes.