5.4.2

LF (Line Feeds) Number Entry (A9)

1. While in the User Menu mode, scroll to "

A 9

", then scroll down once using the 6 key

to enter the Line Feeds menu.

2. The display will momentarily show "

ET LF

", followed by the current line feeds value.

3. Use the front panel numeric keys to key-in the actual line feeds value. If you make a

mistake, press the CLEAR key to clear your entry and start over.

4. After entering the exact value, press the SAMPLE / SAVE key to save the line feeds

value. The display will show "

END LF

" momentarily, then revert back up to A9.

CHAPTER 6: CALIBRATION

6.1

CALIBRATION OVERVIEW

Your S610 scale ships from the factory fully calibrated. There is no need to calibrate the scale

unless you feel that it has become inaccurate. It is a good idea to check the calibration of your scale

from time to time with a precision test weight. Salter Brecknell recommends that you perform a new

calibration on your digital scale at least once a year.

During calibration, at least two values are saved in the scale’s memory – the zero value

(deadweight) and the up to three span values (test weights). See Table 6-1 for a listing of

minimum and recommend test weights.

6.2

CALIBRATION MODE

1. Toggle the ON/OFF switch to the OFF (0) position.
2. Press and hold the ZERO key while powering back on the scale. When the scale shows

” o Zone” you are in Calibration mode and you may release the ZERO key.

3. Press the SAMPLE / SAVE key. The scale shows briefly shows “C 0”, then displays a value

which remains on the screen. This value is for troubleshooting purposes only. Allow a 20-minute

warm-up period for the load cell and electronic components to become thermally stable.

4. Remove all items from the scale’s platter. Press ZERO to zero the value.
5. Press the SAMPLE / SAVE key to save the zero point value.

The scale shows "0.000" or

"0.0000" depending on scale capacity.

6. Use the numeric keys to enter the test weight value. Please note that the decimal point is fixed

so that if you are entering, for example, 2 lb, you need to type “2” followed by zeroes until the “2”

is to the left of the decimal point.

7. Place the test weight(s) onto the scale’s platter, then press the SAMPLE / SAVE key. If the

calibration was successful, the display will show "END C1" momentarily, then automatically

enter Normal Operating Mode by performing a countdown test. Remove the test weight from the

platter.

NOTE: