off by pressing the TARE key. The scale switches to TARE mode and sets the displayed

weight to zero.

2.

Press the SAMPLE / SAVE key. Press the SAMPLE key. The scale displays “10” and

is prompting you to place ten identical items on the platform.

NOTE:

If you wish to change the sample number, simply press the SAMPLE / SAVE

key repeatedly until the desired sample number appears. Available choices are 5, 10,

25, 50 and 100.

3.

Place the sample items on the platform all at once.

4.

Press the PRINT key. If the sample meets the limits shown in Table 3-2, the scale will

now display the number of pieces on the scale. If it does not, the scale flashes the

UNIT WT/MODE annunciator but still allows piece counting. If this occurs, you should

use a higher sample amount to achieve better piece count accuracy. Please see Table

3-1 for an explanation of the error annunciators.

NOTE1 :

If the scale displays a “Error 10” message, the unit weight of the items you

wish to count is too light for your scale to process at all.

3.3.6 CLEARING THE PIECE COUNT

1. To clear the piece count, either press the CLEAR key to erase the sample or repeat the

steps in Section 3.3.5.

MODEL

Capacity / Graduation

Minimum Piece Weight

Minimum Sample Weight

S610-2

2 lb x 0.0002 lb

0.00016 lb

0.003 lb

S610-5

5 lb x 0.0005 lb

0.0004 lb

0.0075 lb

S610-10

10 lb x 0.001 lb

0.0008 lb

0.015 lb

S610-20

20 lb x 0.002 lb

0.0016 lb

0.03 lb

S610-1M

1000 g x 0.1 g

0.08 g

1.5 g

S610-2M

2000 g x 0.2 g

0.16 g

3.0 g

S610-5M

5000 g x 0.5 g

0.4 g

7.5 g

S610-10M

10,000 g x 1 g

0.8 g

15.0 g

TABLE 3-2: S610 Sampling Limits