B140 Counting Scale User Instructions

5.6 Set Scale’s ID

Set the scale’s ID by following these instructions:

1.

In normal working mode, hold the ON/OFF and PLU6 keys at the same time
until …

SEtUP is displayed in the PIECE WEIGHT window.

id (ID) is displayed in the COUNT window.

xxxxxx is displayed in the WEIGHT window. This means that the scale’s
ID is currently set to xxxxx.

2.

Use the numeric keys to change the scale’s ID and press the Enter key save
the setting …

The scale returns to normal mode.

5.7 Displayed Messages

Below are the screen messages you may see during operation of this scale and what
they mean.

Err01

Scale overloaded

– – – – – –

Too many accumulations or accumulated counts too high

Err03

Weight signal is too small

Err04

The zero point is too high

Err05

The zero point is too low

Err12

The setting parameter(s) is not in normal range, or the parameter
has not been set

CAL.Err

There is an error in calibration: maybe the loaded weight is too
small or the input weight is not correct.

Err30

Input signal of ADC is over max. range

Err31

Input signal of ADC is below min. range

CAP

Displaying data relating to capacity

UoL

Displaying data relating to voltage

Add

The accumulated data

PC.t

Data about piece weight

SPL.PCS

Data about sample pieces

uP.PCS

Data about upper limit pieces

Lo.PCS

Data about lower limit pieces

unLd

Unload weight on the platter

InP.Ld

Input loaded weight’s number

CAL.oN

Calibration enable switch is ON

CAL.oFF

Calibration enable switch is OFF

ERR01

Displayed data is over 999999 or down -99999 (not include
decimal point)