4rear view of front face panel, Rear view of front face panel – Cooper Instruments & Systems DPG 100 Digital Pressure Test Gage User Manual
Page 13
Apply the indicated pressure to the instrument. Press [Clear] until "-P-" is
displayed, indicating that the reading is being stored.
Finally, the display will begin to alternately flash between the pressure required for calibration point #2 (for example,
"100.0”) and ”- ---".
Apply the indicated pressure to the instrument. Press [Clear] until "-P-" is
displayed, indicating that the reading is being stored.
When the last pressure point has been entered, the instrument will turn itself off. At this time, the mode jumper must
be moved from its "calibration" position to its "park" position as indicated in Figure 4-1. Close the unit up and
replace the center screw.
Check that the instrument has been calibrated properly by turning it back on and using the pressure source.
4.4
Rear View of Front Face Panel
Figure 4-1: Front face panel (rear view), showing mode jumper positions and orientation of sensor cable
DPG 100
10
Rev Feb 2001