Bio-Rad VersaFluor™ Fluorometer User Manual
Page 25

Section 8
Filters for Fluorescent Protocols
Excitation
Catalog
Emission
Catalog
Application
Filters
Number
Filters
Number
DNA Quant (Hoechst 33258)
EX 360/40 (340–380 nm)
170-2420
EM 460/10 (455–465 nm)
170-2421
ß-Galactosidase (MUG)
EX 360/40 (340–380 nm)
170-2420
EM 460/10 (455–465 nm)
170-2421
ß-Glucuronidase (MUGluc)
EX 360/40 (340–380 nm)
170-2420
EM 460/10 (455–465 nm)
170-2421
Apopain (Z-DEVD-AFC)
EX 360/40 (340–380 nm)
170-2420
EM 520/10 (515–525 nm)
170-2424
pH indicator (BCECF)
EX 490/10 (485–495 nm)
170-2422
EM 520/10 (515–525 nm)
170-2424
Calcium indicator (Fluo-3)
EX 490/10 (485–495 nm)
170-2422
EM 520/10 (515–525 nm)
170-2424
DNA Quant (PicoGreen™)
EX 480/20 (470–490 nm)
170-2427
EM 520/10 (515–525 nm)
170-2424
Oligo Quant (OliGreen™)
EX 480/20 (470–490 nm)
170-2427
EM 520/10 (515–525 nm)
170-2424
Protein Quant (NanoOrange™)
EX 480/20 (470–490 nm)
170-2427
EM 590/10 (585–595 nm)
170-2425
Fluorescein
EX 490/10 (485–495 nm)
170-2422
EM 520/10 (515–525 nm)
170-2424
Ethidium Bromide
EX 510/10 (505–515 nm)
170-2423
EM 620/10 (615–625 nm)
170-2426
RNA Quant (RiboGreen™)
EX 490/10 (485–495 nm)
170-2422
EM 520/10 (515–525 nm)
170-2424
Section 9
Specifications
Instrument
Light source
20 watts incandescent quartz halogen 350 nm to 900 nm.
The quartz halogen lamp is rated to operate for 2000 hours.
Sensing device
Photodiode
Optical filters
Excitation
1” diameter with holder
Emission
1” diameter with holder
Sample source
Cuvette: 12.5 mm x 12.5 mm (outside dimension)
Sensitivity selection
Three settings: LOW, MED, and HIGH.
Approximate factor of 10 between each setting.
Dynamic range
Greater than 5 orders using all three sensitivity selec-
tions.
Printer output
RS-232, Serial
AC Power Requirement
AC power input
100–240 VAC 50–60 Hz, 1.6 A slow-blow fuse
Size and Weight
Overall size
23 cm (L) x 33 cm (W) x 15 cm (H)
Shipping weight
4 kg
Environmental Requirements
Storage environment
0 to 70 °C, humidity 0–95% (non-condensing)
Operating environment
20 to 35 °C, humidity 0–95%
Regulatory
Meets the requirements of:
EN61010-1, EN50082-1, EN55011/03