beautypg.com

2 instructions for use – Bio-Rad Bio-Gel P Polyacrylamide Gel User Manual

Page 11

background image

Table 4. Formulations for buffers and solutions provided in the GST

kits.

Profinia Solution

Supplied Concentrated Formulation

Working 1x Formulation

Volume Position #

As

GST lysis buffer

2x

300 mM NaCl, 20 mM Na

2

HP0

4

,

150 mM NaCl, 10 mM Na

2

HP0

4

, 125 ml

N/A

10 mM EDTA, pH 7.4

5 mM EDTA, pH 7.4

GST wash buffer

2x

300 mM NaCl, 20 mM Na

2

HP0

4

,

150 mM NaCl, 10 mM Na

2

HP0

4

, 200 ml

1

10 mM EDTA, pH 7.4

5 mM EDTA, pH 7.4

GST elution buffer

2x

40 mM glutathione, 200 mM Tris, 20 mM glutathione, 100 mM

100 ml

3

10 mM EDTA, pH 8.0

Tris, 5 mM EDTA, pH 8.0

Desalting buffer

5x

685 mM NaCl, 13.5 mM KCl,

137 mM NaCl, 2.7 mM KCl,

200 ml

4

21.5 mM Na

2

HP0

4

, 40.5 mM

4.3 mM Na

2

HP0

4

,

KH

2

HP0

4

, pH 7.0

8.1 mM KH

2

P0

4

, pH 7.4

(pH 7.4 upon dilution)

Cleaning solution 1

2x

1,000 mM NaCl, 100 mM Tris,

500 mM NaCl, 50 mM Tris,

125 ml

5

pH 8.0

pH 8.0

Cleaning solution 2

4x

2,000 mM NaCl, 400 mM NaOAc, 500 mM NaCl, 100 mM NaOAc,

125 ml

6

pH 4.5

pH 4.5

Storage solution

2x

4% C

6

H

5

CH

2

OH 2%

C

6

H

5

CH

2

OH 200

ml

7

(benzyl alcohol)

(benzyl alcohol)

4.2 Instructions for Use

The buffers are provided as concentrates and are diluted to a

1x formulation by the instrument using the preprogrammed methods. The
entire contents (1.23 g) of the glutathione pack need to be added to the
GST elution buffer prior to a purification. Add 5 ml of the elution buffer to
the vial of glutathione and mix to completely resuspend. Either pipet or
pour the glutathione solution into the elution buffer bottle. Filter the solution
through a 0.2 µm filter to remove particulates. The position labeling of the
bottles and insertion into the instrument is identical to that described in
Section 2.2. Note: GST methods do not use a second wash buffer
and position 2 is left empty or filled with water

. Refer to Figure 4 for

proper placement of the GST buffer bottles. If less than 10 ml of purifications
are performed, the glutathione-containing elution buffer can be frozen in
aliquots and used within 3 months. The stored elution buffer should only
be thawed once. The other GST buffers can be capped and stored at 4°C
for up to 3 months.

9

10006044D:4006073B.QXD

11/4/2009

1:28 PM

Page 9

This manual is related to the following products: