beautypg.com

Assay workflow – Bio-Rad Human MMP and TIMP Assays User Manual

Page 8

background image

6

Prewet wells

(for filter plate only)

Add 50 µl 1x beads to wells

Wash 2 x 100 μl

Add 50 μl diluted standards, samples, controls,

incubate 1 hr at RT with shaking at 850 rpm

Wash 3 x 100 μl

Add 25 μl 1x detection antibody, incubate

30 min at RT with shaking at 850 rpm

Wash 3 x 100 μl

Add 50 μl 1x streptavidin-PE, incubate

10 min at RT with shaking at 850 rpm

Wash 3 x 100 μl

Resuspend in 125 μl assay buffer,

shake at 850 rpm for 30 sec

Acquire data on Bio-Plex system

Assay Workflow