Prepare wash method, Prepare wash method 10, Setting up the bio-plex pro wash station – Bio-Rad Human MMP and TIMP Assays User Manual
Page 12: Setting up the bio-plex handheld magnetic washer
10
2. Select OK and follow the software prompts for step-by-step
instructions for CAL1 and CAL2 calibration.
Note:
In Bio-Plex Manager version 6.1 and higher, startup, warm up,
and calibration can be performed together by selecting the “Start up and
calibrate” icon.
3. Prepare Wash Method
Bio-Plex Pro assays are compatible with both magnetic separation and
vacuum filtration methods. However, for best results, we recommend
performing the assays in a flat bottom plate with magnetic separation.
Setting Up the Bio-Plex Pro Wash Station
The wash station does not require calibration; however, it should be primed
before use. For more information, refer to the Bio-Plex Pro wash station
quick guide (bulletin #5826).
1. Install the appropriate plate carrier on the wash station.
2. Use the prime procedure to prime channel 1 with wash buffer.
Setting Up the Bio-Plex Handheld Magnetic Washer
Place an empty flat bottom plate on the magnetic washer by sliding
it under the retaining clips. Push the clips inward to secure the plate.
Make sure the plate is held securely. If needed, the clips can be adjusted
for height and tension. For detailed instructions, refer to the user guide
(bulletin #10023087).
Table 3. Summary of compatible wash stations and plate types.
Wash Method
Wash Station
Assay Plate
Magnetic separation
Bio-Plex Pro
™
Flat bottom plate
Bio-Plex handheld magnetic washer
Vacuum filtration
Vacuum manifold (manual)
Filter plate