beautypg.com

Bio-Rad Human MMP and TIMP Assays User Manual

Page 32

background image

30

Possible Causes

Low Bead Count

Miscalculation of bead dilution

Beads clumped in multiplex
bead stock tube

Vacuum on for too long when
aspirating buffer from wells


Assay plate not shaken enough
during incubation steps and prior
to reading


Reader is clogged

Incorrect needle height of the
reader

Low Signal or Poor Sensitivity

Standards reconstituted incorrectly

Detection antibody or

streptavidin-PE diluted incorrectly

Possible Solutions

Check your calculations and be
careful to add the correct volumes.

Vortex for 30 sec at medium speed
before aliquoting beads.

Do not apply vacuum to the filter
plate for longer than 10 sec after the
buffer is completely drained from
each well.

Shake the plate at 850 ± 50 rpm
during incubation steps and for
30 sec immediately before reading
the plate.

Refer to the troubleshooting guide
in the Bio-Plex

®

system hardware

instruction manual (bulletin
#10005042).

Adjust the needle height to coincide
with the plate type provided in the kit.

Follow the instructions carefully.

Check your calculations and be
careful to add the correct volumes.