beautypg.com

Bio-Rad Human MMP and TIMP Assays User Manual

Page 33

background image

31

Possible Causes

High Background Signal

Incorrect buffer was used
(for example, assay buffer used
to dilute standards)

Accidentally spiked blank wells


Detection antibodies or
streptavidin-PE incubated too long

Poor Recovery

Expired Bio-Plex reagents
were used


Incorrect amounts of components
were added

Microplate shaker set to an
incorrect speed

High end saturation of the
standard curve

Controls do not fall within
expected ranges.

Possible Solutions

Use diluent HD to dilute standards.

Do not add any antigens to the
blank wells.

Follow the procedure incubation
time precisely.

Check that reagents have not expired.
Use new or nonexpired components.

Check your calculations and be
careful to add the correct volumes.

Check the microplate shaker speed
and use the recommended setting.
Setting the speed too high may
cause splashing and contamination.
Use the recommended plate shaker.

Make sure that correct shaker speed
and incubation times are used. Remove
S1 from data analysis if needed.

Make sure that the vial of controls
is reconstituted at the same time as
standards and in the same diluent.
Incubate for precisely 30 min.