Problem
Cause
Solution
4. Poorly defined fraction-
a.
Protein bands are
a. Carefully align the
ation of proteins.
not parallel to the
cutting template with the
elution chambers.
dye front and molecular
markers of the gel when
cutting.
5. Running conditions
a.
Buffer concentration
a. Make fresh buffer.
outside recommended
or pH is incorrect.
range.
6. Splattering of sample
a.
Vacuum is too strong.
a. Decrease vacuum.
in vacuum harvester.
7. Harvester is collecting
a.
Vacuum is too strong.
a. Decrease vacuum.
uneven sample volumes.
Section 9
Equipment and Accessories
Catalog
Number
Product Description
165-1250
Whole Gel Eluter
165-1251
Whole Gel Eluter with large harvester
165-1255
Mini Whole Gel Eluter
165-1256
Mini Whole Gel Eluter with Mini Harvester
165-1260
Large Harvester
165-1261
Mini Harvester
Consumables
165-1270
Whole Gel Eluter Template
165-1271
Mini Whole Gel Eluter Template
165-1275
Cellophane for Whole Gel Eluter
, 25
165-1276
Cellophane for Mini Whole Gel Eluter
, 25
165-1277
Sealing Tabs for Whole Gel Eluter
, 50
165-1278
Sealing tabs for Mini Whole Gel Eluter
, 50
165-1279
Roller
, 1
165-1280
Lower Chamber filter paper for Whole Gel Eluter
, 75
165-1281
Upper Chamber filter paper for Whole Gel Eluter
, 50
165-1282
Lower Chamber filter paper for Mini Whole Gel Eluter
, 50
165-1283
Upper Chamber filter paper for Mini Whole Gel Eluter
, 50
223-9563
DPTP, non-sterile, style M
, 500
223-9523
DPTP, non-sterile, style D
, 500
223-9645
Culture tubes, 12 x 75 mm
, 5 ml capacity, non-sterile, natural
polypropylene, 2,000
223-9500
Micro Test Tube, 1.5 ml,
polypropylene, natural, capless, 500
22