beautypg.com

Bio-Rad Model 225 Tube Gel Casting Stand User Manual

Page 9

background image

Section 4
Safety Instructions

Power to the Mini-PROTEAN tube cell is supplied by an external DC voltage

power supply. This power supply must be ground isolated in such a way that

the DC voltage output floats with respect to ground. All of Bio-Rad's power

supplies meet this important safety requirement. Regardless of which power

supply is used, the maximum specified operating parameters for the cell are:

1500 VDC

maximum voltage limit

5 Watts

maximum power limit

50 °C

maximum ambient temperature limit

Current to the cell, provided from the external power supply, enters the unit

through the lid assembly, providing a safety interlock to the user. Current to the cell

is broken when the lid is removed. Do not attempt to circumvent this safety

interlock, and always turn the power supply off before removing the lid, or

when working with the cell in any way.

Since potentially dangerous voltages may be used in the electrophoresis cell, it

is important that certain precautions are observed for the safety of the operator.

Please become familiar with these precautions, both for safe operation of this

equipment and to prevent the loss of valuable experimental results.

1. Always turn off the power supply before removing the safety cover of the

electrophoresis cell. Do not use the safety cover as an on-off switch.

2. Never attempt to operate the electrophoresis cell without the safety cover.

3. Do not attempt to move the electrophoresis cell with the power on.

4. Do not leave an operating electrophoresis cell unattended for long periods.

Overnight runs at high voltages are to be avoided. Overnight runs at voltages up to

400 V are acceptable.

5. Do not overfill the upper or lower buffer reservoirs. This can cause current leaks

and arcing. Refer to Section 6.1 for proper buffer levels.

6. Insure that the upper horizontal surfaces of the tube gel module are dry before

starting IEF separation, especially when high voltages are used.

7. If the electrophoresis cell appears to be drawing excessive current, or if the current

fluctuates, turn off the power supply and check the entire system (for example,

buffer chamber or sample reservoir leaks, overfilled buffer chamber, buffer too con-

ductive, etc.).

8. Never allow untrained personnel to use this equipment without proper supervision.

6

!

!

!

This manual is related to the following products: