Silver stain troubleshooting, Sypro ruby troubleshooting – Bio-Rad High-Throughput Dodeca™ Gel Stainers User Manual
Page 22

Silver Stain Troubleshooting
Problem
Cause
Solution
7. Dark background
a. Low quality reagents
a. Contaminants will
or contaminated
catalyze non-specific
water
formation of metallic
silver. Use high purity
reagents and deionized
water
b. Over development
b. Monitor the development
of the gel
step carefully and stop
quickly by draining
through both drain ports
(stopcocks). Then
quickly add the stop
solution through the
reagent access door of
the lid. The drain/fill
cycle should be no
longer than 3 minutes
c. Overloaded protein
c. Adjust the protein sample
sample
load so it is appropriate
for this detection
method
d. Incorrect preparation
d. Refer to the silver stain
of silver stain
instruction manual,
solutions
paying special attention
to the image developer
concentrate amounts
8. Poor sensitivity
a. Inactive reagents or
a. Use the silver stain kit
inaccurate dilution or
components within their
incorrect incubation
recommended shelf life
times
Confirm solution
preparation and
incubation times are
correct. Pay special
attention to the amount
of image developer
concentrate
b. Low protein sample
b. Adjust the protein sample
load
load so it is appropriate
for this detection
method
Sypro Ruby Troubleshooting
Problem
Cause
Solution
9. Weak or no
a. Residual Coomassie
a. Clean the Dodeca stainer
fluorescent signal
stain in the system
thoroughly before
inhibits the fluorescent
switching from
signal
Coomassie based stain
to fluorescent stains
b. Low protein sample
b. Adjust the protein sample
load
load so it is appropriate
for this detection method
19