Problem
Cause
Solution
9. End gels are running
a. Buffer is warmer at the
a. Ensure the stir bar is freely
faster than the center
ends of the cell than the
rotating to maintain a
gels.
center.
constant buffer temperature.
10. Sample not migrating
a. Tape is still on the bottom
a. Remove the white tape
of the gel.
prior to electrophoresis.
b. No electrical connection.
b.Check continuity of
electrode wires in lid and
tank.