4b. An alternative method is to run the comb down the sides of the cassette to break the weld-
joint and to pull the halves apart.
Fig. 7. Alternative method for opening Criterion Gel cassette after electrophoresis.
5. It is best to remove the gel by floating it off the plate. Invert the gel and plate under fixative
or transfer solution and agitate gently until the gel separates from the plate.
6. Rinse the Criterion tank and lid with distilled, deionized water after use.
Section 3
Maintenance
C
Crriitte
erriio
on
n tta
an
nk
k a
an
nd
d lliid
d
Rinse thoroughly with distilled water after every
use.
Section 4
Troubleshooting Guide
Problem
Cause
Solution
1. "Smile effect" - band pattern
a.
Center of the gel running
a.
Buffer not mixed well or buffer
curves upward at both sides
hotter than either end.
in upper chamber too
of the gel.
concentrated. Remake buffer,
insuring thorough mixing,
especially when diluting 5x or
10x stock.
b.
Power conditions excessive.
b.
Decrease power setting from
200 V to 150 V or fill lower
chamber to the fill line on the
tank.
6
- TransFectin™ Lipid Reagent (2 pages)
- Gene Pulser MXcell™ Electroporation System (19 pages)
- Gene Pulser MXcell™ Electroporation System (66 pages)
- Gene Pulser Xcell™ Electroporation Systems (83 pages)
- Gene Pulser® Electroporation Buffer (2 pages)
- MicroPulser™ Electroporator (31 pages)
- Helios® Gene Gun System (52 pages)
- PDS-1000 / He™ and Hepta™ Systems (51 pages)
- TGX™ FastCast™ Acrylamide Solutions (2 pages)
- Criterion™ TGX Stain-Free™ Precast Gels (16 pages)
- Image Lab™ Software (212 pages)
- Gel Doc™ EZ System (22 pages)
- Criterion Stain Free™ Tris-HCl Gels (96 pages)
- Mini-PROTEAN® TGX™ Precast Gels (52 pages)
- Image Lab™ Software (236 pages)
- ChemiDoc™ XRS+ System (42 pages)
- ChemiDoc™ XRS+ System (4 pages)
- ChemiDoc™ XRS+ System (50 pages)
- GS-800™ Calibrated Densitometer (444 pages)
- ChemiDoc™ MP System (8 pages)
- Criterion™ TGX™ Precast Gels (60 pages)
- Image Lab™ Software (260 pages)
- Criterion™ XT Bis-Tris Precast Gels (26 pages)
- 2-D Electrophoresis Workflow (14 pages)
- Fluorescent Protein Stains (27 pages)
- 2-D Electrophoresis Workflow (22 pages)
- 2-D Electrophoresis Workflow (20 pages)
- PROTEAN® i12™ IEF System (4 pages)
- ReadyPrep™ 2-D Starter Kit (28 pages)
- EXQuest Spot Cutter (81 pages)
- Ready Gel® Zymogram Precast Gels (46 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (44 pages)
- Precision Plus Protein™ Prestained Standards (3 pages)
- Precision Plus Protein™ Unstained Standards (16 pages)
- Prestained SDS-PAGE Standards (3 pages)
- Unstained SDS-PAGE Standards (3 pages)
- Silver Stains (20 pages)
- Biotinylated Standards (3 pages)
- Biotinylated Standards (11 pages)
- IEF and 2-D Standards (3 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (20 pages)
- Mini-PROTEAN 3 Multi-Casting Chamber (10 pages)
- PROTEAN® Plus Hinged Spacer Plates and Combs (21 pages)
- Mini-PROTEAN® Tetra Handcast Systems (10 pages)