1.3 Safety

Power to the Criterion cell is supplied by an external DC voltage power supply (not included).

The output of this power supply must be isolated from external ground to insure that the DC
voltage output floats with respect to ground. All Bio-Rad power supplies meet this
important safety requirement. Regardless of the power supply used, the maximum specified
operating parameters for the Criterion Cell are as follows:

600 VDC

maximum voltage limit

15 watts

maximum power limit

50 °C

maximum ambient temperature limit

The current to the cell enters the unit through the lid assembly that provides a safety

interlock to the user. The current to the cell and the lid's upper electrodes is broken when the
lid is removed. Always turn off the power supply before removing the lid. Do not attempt to use
the cell without the safety lid.

Important: This Bio-Rad product is designed and certified to meet *EN61010-1 safety
standards. Certified products are safe to use when operated in accordance with the instruction
manual. This instrument should not be modified or altered in any way. Alteration of this
instrument will

Void the warranty

Void the EN61010-1 certification, and

Create a potential safety hazard

Bio-Rad is not responsible for any injury or damage caused by use of this instrument for

purposes other than those for which it is intended or by modifications of the instrument not
performed by Bio-Rad or an authorized agent.

*EN61010-1 is an internationally accepted electrical safety standard for laboratory instruments.

Section 2
Set Up and Basic Operation

Criterion precast gel Cassette Preparation (see Criterion precast gel instruction manual for

details)

1. Remove the Criterion gel cassette from the storage container.

2. Gently remove the comb and rinse the wells thoroughly with distilled water or running

buffer.

3. Remove the tape from the bottom of the cassette.

4. Repeat for second Criterion gel.