Amico NFPA Triplex RVL Modular Stack Mount User Manual
Page 42
Amico Source Equipment
2
Control Panel
Vacuum System
Accessories
Vacuum Pump
Air Receiver
• UL listed control panel in a NEMA 12 enclosure with a 5.7”
Touch Screen Display
• Only panel components that are commercially available and
not of propriety design will be considered
• Our pumps include the following accessories:
- Temperature sensor at each discharge port of the pump
with numerical display
- Externally operable circuit breakers with door interlock
per pump
- Control circuit transformers with fused primary and
secondary circuit breakers
- WinCE 5.0 operating system
- Illuminated H-O-A switch
- Magnetic starter with three leg overload protection
- Hour meter and motor running light
• Audible and visual local alarms with dry contacts for the
following:
- Lag pump in use
- High discharge air temperature
- Main transformer failure
- Motor overload
• Provide manual reset for all alarms
• All control and alarm functions shall remain energized while
any pump in the system remains electrically on-line
• Ethernet connectivity with embedded web-page for remote
monitoring, password protected remote adjustment of all
operating parameters
• Lag pump shall be able to start automatically if the lead
pump fails to operate
• Alarm contacts will be provided for remote annunciation for
all alarm points
• Shutdowns for systems are as per NFPA
• Vacuum pump shall be of the continuous
duty, high efficiency, Oil-Less and frictionless
compression Contact-Less Claw type
• Pumping chamber is oil free
• Pump is completely air cooled with no water
requirements
• Vacuum pump shall be driven through a
single lubricated gear-box
• Torque is transmitted from the motor to the
pump through a shaft coupling
• Claws are synchronized by heavy-duty
precision timing gears
• Each vacuum pump shall have a factory
piped intake with integral flex connector,
isolation valve and check valve
• Each pump shall have a built-in anti-suck
back valve mounted at the pump inlet
• System should also have a means of
removal of the vacuum pump for service
or replacement without interruption to the
system
• The pump shall include:
- An inlet air filter (optional)
- Check valve
- Flexible connectors
- Isolation valve
- High discharge temperature
sensor (with visual indicator)
- Drain valve
- Numerical temperature display
- Site glass
• Shall be ASME certified with provisions for isolating the
receiver without interruption of the vacuum service
• Receiver shall be rated for full vacuum service with a minimum
200 psig design pressure
• Shall be equipped with a vacuum gauge, manual drain valve,
and a shut-off valve
• The inside of the receiver consists of a two-part epoxy coating
providing rust protection equal or better than that achieved by galvanizing
Amico Source | 85 Fulton Way, Richmond Hill, ON, L4B 2N4, Canada Toll Free Tel: 1.877.462.6426 | Tel: 905.764.0800 | Fax: 905.764.0862
www.amico.com
40
- NFPA Triplex RVD Modular Stack Mount NFPA Triplex CCD Modular Stack Mount NFPA Quadruplex RVL Modular Stack Mount NFPA Quadruplex RVD Modular Stack Mount NFPA Quadruplex CCD Modular Stack Mount NFPA Duplex RVL Vertical Tank Mount NFPA Duplex RVL Modular Stack Mount NFPA Duplex RVL Horizontal Tank Mount NFPA Duplex RVD Vertical Tank Mount NFPA Duplex RVD Modular Stack Mount NFPA Duplex RVD Horizontal Tank Mount NFPA Duplex CCD Vertical Tank Mount NFPA Duplex CCD Modular Stack Mount NFPA Duplex CCD Horizontal Tank Mount