Amico Source Equipment

2

Control Panel

Vacuum System
Accessories

Vacuum Pump

Air Receiver

• UL listed digital control panel in a NEMA 12 enclosure with a
5.7” Touch Screen Display
• Our pumps include the following accessories:
- Temperature sensor at each discharge port of the pump
with numerical display
- Externally operable circuit breakers with door interlock
per pump
- Control circuit transformers with fused primary and
secondary circuit breakers
- WinCE 5.0 operating system
- Illuminated H-O-A switch
- Magnetic starter with three leg overload protection
- Hour meter and motor running light
• Lag pump shall be able to start automatically if the lead
pump fails to operate

• Audible and visual local alarms with dry contacts for the
following:
- Lag pump in use
- High discharge air temperature
- Main transformer failure
- Low oil level
- Motor overload
• Provide manual reset for all alarms
• All control and alarm functions shall remain energized while
any pump in the system remains electrically on-line
• Ethernet connectivity with embedded web-page for remote
monitoring, password protected remote adjustment of all
operating parameters
• Alarm contacts will be provided for remote annunciation for
all alarm points
• Shutdowns for systems are as per NFPA

• Vacuum pump shall be an Oil-Sealed Rotary Vane type

with air cooled design
• Dynamically balanced multi-vane design
• Oil lubrication system shall be a pressure differential

design with full re-circulation and multi-stage exhaust oil
separation rated at 99.998% efficiency
• Oil cooler shall be a single pass, cross flow type with a

dedicated cooling fan
• Bearings shall be permanently lubricated and sealed
• Water vapor condensation in the cylinder shall be

prevented by means of an automatic gas ballast valve

• Vacuum pump shall be direct driven
• Torque is transmitted from the motor to the pump

through a shaft coupling
• Each vacuum pump shall be factory piped intake with

integral flex connector, isolation valve and check valve
• Each pump shall have a built-in anti-suck back valve

mounted at the pump inlet
• System should also have a means of removal of the

vacuum pump for service or replacement without

interruption to the system

• The pump shall include:
- Non-return valve
- Back pressure gauge
- High discharge temperature
sensor (with visual indicator)
- Oil drain valve
- Oil sight glass
- Flexible connector
- Isolation valve
- Check valve

• Shall be ASME certified with provisions for isolating the

receiver without interruption of the vacuum service
• Receiver shall be rated for full vacuum service with a minimum
200 psig design pressure
• Shall be equipped with a vacuum gauge, manual drain valve,
and a shut-off valve
• The inside of the receiver consists of a two-part epoxy coating
providing rust protection equal or better than that achieved by galvanizing