Calibration, Setting the tds factor – LaMotte AT Visual Series Water Softener Demo Kit User Manual
Page 17

17
TDS • CODE 5-0080 • Dual Range TDS Pocketester
BEFORE FIRST USE: Soak electrodes for a few minutes in alcohol to remove oils.
CALIBRATION
Use a calibration standard between 200 ppm and 1,900 ppm for low range testing. 
Use a calibration standard between 1,900 ppm and 7,000 pm for high range testing.
See page 20 for suggested standards.
SETTING
THE
TDS
FACTOR
This tester lets you select a TDS factor of 0.4 to 1.0. The preset TDS factor of 0.7 is recommended.
1. Open battery compartment. With meter on, press HOLD key, then press the INC key. (INC key is inside 
battery compartment).
2. Press the INC or DEC keys to adjust the TDS factor.
3. After 3 seconds without a key press, the display fl ashes 3 times, then shows “ENT”. Tester accepts TDS 
factor and returns to measurement mode.
4. Replace battery cap.
(To Be Done Once a Month)
Pour ½” to 1” 
of a Calibration 
Standard into 2 
containers, and 
tap or deionized 
water into a third 
container.
Open battery compartment lid (end 
with lanyard loop). The two white 
buttons are Increment (INC) and 
Decrement (DEC) calibration keys.
 Rinse electrode
 in water. Then 
rinse in fi rst container of 
standard, then dip into 
the second.
Press in INC or DEC keys to 
adjust reading to match the 
calibration standard value. 
After 3 seconds without a key press, 
the display fl ashes 3 times, then 
shows “ENT”. The tester 
accepts calibration value; returns to 
measurements mode.
Replace battery cap. 
Rinse the electrode in the 
water and proceed with 
test procedure.
1.
2.
3.
4.
5.
6.
 Turn the meter on, and wait 
several minutes 
to allow the 
display to stabilize.