beautypg.com

LaMotte WaterLink 3 Express Lab User Manual

Page 14

background image

PHOSPHATE PROCEDURE
Remove the UDV Adapter (1724) from the meter chamber. (See page 8)

Press and briefly hold

1.

to turn the meter on.

Press

2.

to scroll to Other Series.

Press

3.

ENTER

to select Other Series.

Press

4.

to scroll to Phosphate.

Press

5.

ENTER

to select Phosphate.

Rinse a clean, glass test tube (0290) with sample water. Fill to the 10

6.

mL line with the sample.
Insert tube into chamber, close lid and press

7.

ENTER

to select BLANK

vial.

NOTE: If a blank has been scanned previously in the same type of

tube, the meter will not require that a blank be scanned.
Remove the tube from the colorimeter.

8.

Rinse a clean plastic test tube (0106) with sample water. Fill to the 10

9.

mL line with the sample.
Remove one Phosphate LR Test Strip (3021) from the vial. Gently

10.

bend strip in half (do not fold) with pads facing inward. Place strip

inside test tube cap.
Cap tube and slowly invert 5 times. In one complete inversion, the

11.

tube will be turned cap down and then right side up. The air bubble will

move slowly to the bottom of the tube and back again to the cap end.

Five inversions should take about 15 seconds.
Remove the cap and test strip from the plastic tube (0106). Pour the

12.

reacted sample into a glass tube (0290). Cap.
Immediately insert tube into chamber, close lid and select SCAN

13.

SAMPLE. Record result as ppb Phosphate.
Follow the next test procedure to test the next test factor in the series,

14.

press

or

to select another test in the series, press

EXIT

to

escape to a previous menu, or press

to turn the meter off.

NOTE: The Natural Chemistry

®

PHOS-TEST strip can be used for this test.

Follow the instructions above.

14

Testing with the W

aterLink

®

3 Express Lab