beautypg.com

LaMotte WaterLink 3 Express Lab User Manual

Page 10

background image

WaterLink

®

3 Express COUNTER MAT

Place the colorimeter on the left side of the counter mat, then place both

racks on the mat. Insert the UDVs (Unit Dose Vials) for each test factor into

the corresponding row.

NOTE: The timing sequence for Unit Dose Vials color development on the

counter mats is very important. Note that the test for free chlorine should be

read immediately.

UNIT DOSE VIALS (UDVS)

Each box contains 100 UDVs (20 pouches of 5 each) for one specific

1.

test factor. Remove one pouch from each box for tests to be conducted

over the next 24 hours.
Avoid opening more pouches than you need for one day. Some UDVs

2.

have a limited shelf life and should not be exposed to the humidity in

the air more than necessary. A foil storage bag with a desiccant packet

is included that can extend the shelf life to several days.
Open pouches by tearing them, or cutting them with scissors,

3.

horizontally across the top, then remove the group of 5 UDVs.
Place each set of individual UDVs in the rack directly behind their

4.

respective test factor as noted on the counter mat.

TESTING WITH THE WATERLINK

®

3 EXPRESS LAB

CHLORINE & BROMINE UDV SERIES

Press

1.

briefly to turn the meter on.

Press

2.

or

to scroll to CHLORINE SERIES or BROMINE

SERIES.
Press

3.

ENTER

to select CHLORINE SERIES or BROMINE SERIES.

Insert the pipettor into the pool or spa water sample to be tested.

4.

Slowly extract the pool or spa water sample that will be measured.
The first vial will be the “Blank”. An empty vial containing 3 mL of pool

5.

or spa water sample will be used to clear the colorimeter memory and

prepare it for the series of tests. Press the plunger to dispense the

water sample into the BLANK vial.
Insert the Blank vial into the meter chamber.

6.

Use the pipettor to slowly dispense 3 mL of sample water in each UDV

7.

(Unit Dose Vial) from right to left. If testing the full series, the Borate

UDV will be filled first. If Borate or Nitrate are not to be tested, fill the

Iron UDV with sample water. (The Bromine test can be performed using

Free or Total Chlorine UDVs.)

TIP: Press the pipettor tip through the foil with your finger. Always gently

break the seal of the UDV with the tip of the pipettor and slowly release the

10

Testing with the W

aterLink

®

3 Express Lab