beautypg.com

LaMotte SMART 2 Electronic Soil Lab User Manual

Page 64

background image

PROCEDURE

1. Press and hold ON button until colorimeter turns on.
2. Press ENTER to start.
3. Press ENTER to select TESTING MENU.
4. Select ALL TESTS (or another sequence containing 51 Iron Bipyr)

from TESTING MENU.

5. Scroll to and select 51 Iron Bipyr from menu.
6. Fill a clean tube (0290) to the 10 mL line with the soil extract then

neutralize according to the procedure on page 50.

7. Insert tube into chamber, close lid and select SCAN BLANK.
8. Remove tube from colorimeter. Use the 0.5 mL pipet (0353) to add one

measure of *Iron Reagent #1 (V-4450). Cap and mix.

9. Use the 0.1 g spoon (0699) to add 0.1 g of *Iron Reagent #2 Powder

(V-4451). Cap and shake vigorously for 30 seconds. Wait three minutes for
maximum color development.

10. At the end of 3 minute waiting period, do not mix. Insert tube into

chamber, close lid and select SCAN SAMPLE. Multiply the result by 5 to
determine the iron concentration in ppm.

11. Press OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents are obtained.

n

IRON CONCENTRATION CHART

PARTS PER MILLION

RANGE

0.0-1.3 ppm

Very Low

1.4-3.0 ppm

Low

3.0-5.0 ppm

Medium

5.0-10.0 ppm

Medium High

Above 10.0-25.0 ppm

High

64

SMART2 SOIL MANUAL 1.07