PROCEDURE
1. Press and hold ON button until colorimeter turns on.
2. Press ENTER to start.
3. Press ENTER to select TESTING MENU.
4. Select ALL TESTS (or another sequence containing 51 Iron Bipyr)
from TESTING MENU.
5. Scroll to and select 51 Iron Bipyr from menu.
6. Fill a clean tube (0290) to the 10 mL line with the soil extract then
neutralize according to the procedure on page 50.
7. Insert tube into chamber, close lid and select SCAN BLANK.
8. Remove tube from colorimeter. Use the 0.5 mL pipet (0353) to add one
measure of *Iron Reagent #1 (V-4450). Cap and mix.
9. Use the 0.1 g spoon (0699) to add 0.1 g of *Iron Reagent #2 Powder
(V-4451). Cap and shake vigorously for 30 seconds. Wait three minutes for
maximum color development.
10. At the end of 3 minute waiting period, do not mix. Insert tube into
chamber, close lid and select SCAN SAMPLE. Multiply the result by 5 to
determine the iron concentration in ppm.
11. Press OFF button to turn colorimeter off or press EXIT button to exit to a
previous menu or make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents are obtained.
n
IRON CONCENTRATION CHART
PARTS PER MILLION
RANGE
0.0-1.3 ppm
Very Low
1.4-3.0 ppm
Low
3.0-5.0 ppm
Medium
5.0-10.0 ppm
Medium High
Above 10.0-25.0 ppm
High