SPECIFICATIONS

n

INSTRUMENT TYPE: Colorimeter

Readout

Graphical 4 line, 16 character per line LCD

Wavelengths

430nm, 520 nm, 570 nm, 620 nm

Wavelength Accuracy

± 2 nm

Readable Resolution

Determined by reagent system

Wavelength Bandwidth

10 nm typical

Photometric Range

-2 to +2A

Photometric Precision

± 0.001A

Sample Chamber

Accepts 25 mm diameter flat-bottomed test tubes

Light Sources

4 LEDs

Detectors

4 silicon photodiodes with integrated interference
filters

Modes

Absorbance

, pre-programmed tests

Pre-Programmed Tests

YES, with automatic wavelength selection

RS232 Port

8 pin mini-DIN, 9600b, 8, 1, n

Power Requirements

Battery Operation: 9 volt alkaline
Line Operation: 110/220V AC;
50/60 Hz with adapter, 6V 500 mA DC

Dimensions (LxWxH)

8.5 x 16.2 x 16.7 cm, 3.4 x 6.4 x 2.6 inches

Weight

312 g, 11 oz (meter only)

Data Logger

350 test results stored for download to a PC

6