beautypg.com

LaMotte SMART 2 Electronic Soil Lab User Manual

Page 12

background image

n

SAMPLE DILUTION TECHNIQUES

& VOLUMETRIC MEASUREMENTS

If a test result using the SMART 2 Colorimeter gives an OVERRANGE message
then the sample concentration could be over range or under range. If it is over
range, the sample must be diluted. Then the test should be repeated on the
diluted sample to obtain a reading which is in the concentration range for the
test. (Note: This is not true for colorimetric determination of pH.)
Example:

Measure 5 mL of the water sample into a graduated cylinder. Add
demineralized water until the cylinder is filled to the 10 mL line. The sample
has been diluted by one-half, and the dilution factor is therefore 2. Perform the
test procedure, then multiply the resulting concentration by 2 to obtain the
test result.

The following table gives quick reference guidelines on dilutions of various
proportions. All dilutions are based on a 10 mL volume, so several dilutions
will require small volumes of the water sample. Graduated pipets should be
used for all dilutions.

Size of Sample

Deionized Water to Bring

Volume to 10 mL

Multiplication Factor

10 mL

0 mL

1

5 mL

5 mL

2

2.5 mL

7.5 mL

4

1 mL

9 mL

10

0.5 mL

9.5 mL

20

If the above glassware is not available, dilutions can be made with the
colorimeter tube. Fill the tube to the 10 mL line with the sample then transfer
it to another container. Add 10 mL volumes of demineralized water to the
container and mix. Transfer back 10 mL of the diluted sample to the tube and
follow the test procedure. Continue diluting and testing until a reading, which
is in the concentration range for the test, is obtained. Be sure to multiply the
concentration found by the dilution factor (the number of total 10 mL volumes
used).

Example:

10 mL of sample is diluted with three 10 mL volumes of demineralized water;
the dilution factor is four.

n

INTERFERENCES

LaMotte reagent systems are designed to minimize most common interferences.
Each individual test instruction discusses interferences unique to that test. Be
aware of possible interferences in the soil extract being tested.

The reagent systems also contain buffers to adjust the soil extract to the ideal
pH for the reaction. It is possible that the buffer capacity of the soil extract may
exceed the buffer capacity of the reagent system and the ideal pH will not be
obtained. If this is suspected, measure the pH of a reacted distilled water

12

SMART2 SOIL MANUAL 1.07