The SMART 2 Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press

Q/ENTER to scan

blank.
NOTE: Do not keep the button depressed.

002 Aluminum

* Scan Blank

The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.

002 Aluminum

Blank Done

* Scan Blank

Place the reacted sample in the chamber, close the
lid and press

Q/ENTER to scan sample. The

colorimeter will scan the sample and the results
screen will appear.

002 Aluminum

* Scan Sample

Record test result. To repeat the test, press

Q/ENTER to scan the sample again. The last blank

scanned is used to zero the colorimeter for repeated
scans. A different blank can be used by pressing

s to

scroll back to Scan Blank and then scanning
another blank. Scroll with the

t or s buttons and

make another selection with the

Q/ENTER button.

The %T or Absorbance of the last test can be viewed
by choosing %T/Abs. Press

EXIT to escape to

previous menus.
NOTE: The menus loop in this screen so either the

s or t buttons will lead to the menu selection

needed.

002 Aluminum

0.09 ppm

* Scan Blank

Next Test

Previous Test

%/Abs

Scan Blank