Blank done, 09 ppm – LaMotte SMART 2 Electronic Soil Lab User Manual
Page 25
The SMART 2 Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press
Q/ENTER to scan
blank.
NOTE: Do not keep the button depressed.
002 Aluminum
* Scan Blank
The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.
002 Aluminum
Blank Done
* Scan Blank
Place the reacted sample in the chamber, close the
lid and press
Q/ENTER to scan sample. The
colorimeter will scan the sample and the results
screen will appear.
002 Aluminum
* Scan Sample
Record test result. To repeat the test, press
Q/ENTER to scan the sample again. The last blank
scanned is used to zero the colorimeter for repeated
scans. A different blank can be used by pressing
s to
scroll back to Scan Blank and then scanning
another blank. Scroll with the
t or s buttons and
make another selection with the
Q/ENTER button.
The %T or Absorbance of the last test can be viewed
by choosing %T/Abs. Press
EXIT to escape to
previous menus.
NOTE: The menus loop in this screen so either the
s or t buttons will lead to the menu selection
needed.
002 Aluminum
0.09 ppm
* Scan Blank
Next Test
Previous Test
%/Abs
Scan Blank
SMART2 SOIL MANUAL 1.07